Calculating percent depth dose with the electron pencil‐beam redefinition algorithm

نویسندگان

  • Michael J. Price
  • Kenneth R. Hogstrom
  • John A. Antolak
  • R. Allen White
  • Charles D. Bloch
  • Robert A Boyd
چکیده

In the present work, we investigated the accuracy of the electron pencil-beam redefinition algorithm (PBRA) in calculating central-axis percent depth dose in water for rectangular fields. The PBRA energy correction factor C(E) was determined so that PBRA-calculated percent depth dose best matched the percent depth dose measured in water. The hypothesis tested was that a method can be implemented into the PBRA that will enable the algorithm to calculate central-axis percent depth dose in water at a 100-cm source-to-surface distance (SSD) with an accuracy of 2% or 1-mm distance to agreement for rectangular field sizes > or = 2 x 2 cm. Preliminary investigations showed that C(E), determined using a single percent depth dose for a large field (that is, having side-scatter equilibrium), was insufficient for the PBRA to accurately calculate percent depth dose for all square fields > or = 2 x 2 cm. Therefore, two alternative methods for determining C(E) were investigated. In Method 1, C(E), modeled as a polynomial in energy, was determined by fitting the PBRA calculations to individual rectangular-field percent depth doses. In Method 2, C(E) for square fields, described by a polynomial in both energy and side of square W [that is, C = C(E,W)], was determined by fitting the PBRA calculations to measured percent depth dose for a small number of square fields. Using the function C(E,W), C(E) for other square fields was determined, and C(E) for rectangular field sizes was determined using the geometric mean of C(E) for the two measured square fields of the dimension of the rectangle (square root method). Using both methods, PBRA calculations were evaluated by comparison with measured square-field and derived rectangular-field percent depth doses at 100-cm SSD for the Siemens Primus radiotherapy accelerator equipped with a 25 x 25-cm applicator at 10 MeV and 15 MeV. To improve the fit of C(E) and C(E,W) to the electron component of percent depth dose, it was necessary to modify the PBRA's photon depth dose model to include dose buildup. Results showed that, using both methods, the PBRA was able to predict percent depth dose within criteria for all square and rectangular fields. Results showed that second- or third-order polynomials in energy (Methods 1 and 2) and in field size (Method 2) were typically required. Although the time for dose calculation using Method 1 is approximately twice that using Method 2, we recommend that Method 1 be used for clinical implementation of the PBRA because it is more accurate (most measured depth doses predicted within approximately 1%) and simpler to implement.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Purpose: To evaluate the accuracy of electron dose calculated using the pencil-beam redefinition algorithm (PBRA) in patient-like 2D phantoms. With the recent availability of bolus electron conformal therapy (ECT), conforming of the 90% dose surface to the planning

ID: 16536 Title: Accuracy of electron dose from pencil-beam redefinition algorithm in patient-like 2D phantoms Purpose: To evaluate the accuracy of electron dose calculated using the pencil-beam redefinition algorithm (PBRA) in patient-like 2D phantoms. With the recent availability of bolus electron conformal therapy (ECT), conforming of the 90% dose surface to the planning target volume (PTV) ...

متن کامل

[Electron fields in clinical application. A comparison of pencil beam and Monte Carlo algorithm].

BACKGROUND AND PURPOSE For several years three-dimensional treatment-planning systems have used pencil beam algorithms in the calculation of electron fields. Nowadays, exact Monte Carlo methods are commercially available, showing good correspondence to experimental results. Clinical examples are investigated to find differences in the dose distribution of treatment plans, which are calculated w...

متن کامل

Benchmarking of Monte Carlo model of 6 Mv photon beam produced by Siemens Oncor® linear accelerator: determination of initial electron beam parameters in comparison with measurement

Introduction: The aim of this study was to investigate the initial electron beam parameters for Monte Carlo model of 6MV photon beam produced by Siemens Oncor® linear accelerator.   Materials and Methods: In this study, the EGSnrc Monte Carlo user codes BEAMnrc and DOSXYZnrc were used. The beamnrc code were used for modelling of a 6 MV photon beam produced by...

متن کامل

Dosimetric evaluation of a treatment planning system using pencil beam convolution algorithm for enhanced dynamic wedges with symmetric and asymmetric fields

Background: The dosimetric performance of Eclipse 6.5 three dimensional treatment planning system (3DTPS) is evaluated by comparing the calculated and measured dose in two dimensions following the guide lines of American Association for Physicists in Medicine Task Group 53. Materials and Methods: The calculations were performed by the 3DTPS for symmetric as well as asymmetric fields for standar...

متن کامل

Comparison of MCNP4C, 4B and 4A Monte Carlo codes when calculating electron therapy depth doses

ABSTRACT Background: accurate methods of radiation therapy dose calculation. There are different Monte Carlo codesfor simulation of photons, electrons and the coupled transport of electrons and photons. MCNPis a general purpose Monte Carlo code that can be used for electron, photon and coupledphoton-electron transport.Monte Carlo simulation of radiation transport is considered to be one of the ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 8  شماره 

صفحات  -

تاریخ انتشار 2007