Soft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG Recording
نویسندگان
چکیده
Conventional gel electrodes are widely used for biopotential measurements, despite important drawbacks such as skin irritation, long set-up time and uncomfortable removal. Recently introduced dry electrodes with rigid metal pins overcome most of these problems; however, their rigidity causes discomfort and pain. This paper presents dry electrodes offering high user comfort, since they are fabricated from EPDM rubber containing various additives for optimum conductivity, flexibility and ease of fabrication. The electrode impedance is measured on phantoms and human skin. After optimization of the polymer composition, the skin-electrode impedance is only ~10 times larger than that of gel electrodes. Therefore, these electrodes are directly capable of recording strong biopotential signals such as ECG while for low-amplitude signals such as EEG, the electrodes need to be coupled with an active circuit. EEG recordings using active polymer electrodes connected to a clinical EEG system show very promising results: alpha waves can be clearly observed when subjects close their eyes, and correlation and coherence analyses reveal high similarity between dry and gel electrode signals. Moreover, all subjects reported that our polymer electrodes did not cause discomfort. Hence, the polymer-based dry electrodes are promising alternatives to either rigid dry electrodes or conventional gel electrodes.
منابع مشابه
Conductive Polymer Combined Silk Fiber Bundle for Bioelectrical Signal Recording
Electrode materials for recording biomedical signals, such as electrocardiography (ECG), electroencephalography (EEG) and evoked potentials data, are expected to be soft, hydrophilic and electroconductive to minimize the stress imposed on living tissue, especially during long-term monitoring. We have developed and characterized string-shaped electrodes made from conductive polymer with silk fib...
متن کاملEvaluation of a Low-cost and Low-noise Active Dry Electrode for Long-term Biopotential Recording
Wet Ag/AgCl electrodes, although very popular in clinical diagnosis, are not appropriate for expanding applications of wearable biopotential recording systems which are used repetitively and for a long time. Here, the development of a low-cost and low-noise active dry electrode is presented. The performance of the new electrodes was assessed for recording electrocardiogram (ECG) and electroence...
متن کاملSignal Quality in Dry Electrode EEG and the Relation to Skin-electrode Contact Impedance Magnitude
Current EEG research approaches are focusing on developing new dry electrode EEG (electroencephalogram) systems providing a high enough signal quality for a wide range of applications. This study proposes several parameters for evaluating signal quality of dry electrodes and relates the results to skin-electrode contact impedance magnitude values. The EEG recordings of a Ag/AgCl pinned electrod...
متن کاملCarbon nanofiber-filled conductive silicone elastomers as soft, dry bioelectronic interfaces
Soft and pliable conductive polymer composites hold promise for application as bioelectronic interfaces such as for electroencephalography (EEG). In clinical, laboratory, and real-world EEG there is a desire for dry, soft, and comfortable interfaces to the scalp that are capable of relaying the μV-level scalp potentials to signal processing electronics. A key challenge is that most material app...
متن کاملAn EEG system with carbon wire electrodes and an anti-polarization circuit for simultaneous EEG-fMRI recording
Background Simultaneous EEG-fMRI (Electroencephalography-functional Magnetic Resonance Imaging) recording offers high temporal resolution electrophysiological recording and high spatial resolution hemodynamic recording from the same experimental runs. Carbon wire electrodes (but not solid electrodes with carbon leads) are suitable for simultaneous EEG-fMRI recording, especially in high field MR...
متن کامل