Determination of glucose in human aqueous humor using Raman spectroscopy and designed-solution calibration.
نویسندگان
چکیده
Glucose concentrations of in vitro human aqueous humor (HAH) samples from cataract patients were determined using 785 nm Raman spectra and partial least squares (PLS) calibration. PLS models were created from spectra of prepared calibration solutions rather than aqueous humor samples. Spectra were obtained with an excitation energy (100 mW for 150 s), which was higher than can be applied in vivo, to decrease the models' contribution to prediction uncertainty. The solutions contained experimentally designed levels of glucose, bicarbonate, lactate, urea, and ascorbate. Multiplicative signal correction of spectra helped compensate for the +/-20% drift in laser power observed at the sample over six noncontiguous days of data collection. Seventeen HAH samples containing 38-775 mg/dL of glucose exhibited a root-mean-square error (RMSEP) of 22 mg/dL, coefficient of determination (r(2)) of 0.989, and bias of 6 mg/dL when predicted from lower energy (30 s) spectra collected contemporaneously with fifty calibration spectra. Similar results were obtained even when spectral data were gathered separately for human aqueous humor samples and calibration samples: 10 HAH samples, calibrated on 25 solutions measured 3.6 weeks earlier, exhibited an RMSEP of 23 mg/dL, r(2) of 0.992, and bias of 9 mg/dL. The results demonstrate progress toward the determination of glucose levels in patient-derived aqueous humor using laboratory-derived "artificial aqueous humor" calibration solutions.
منابع مشابه
Non - Invasive Glucose Monitor
Borchert et al., A Noninvasive Glucose Monitor: Preliminary Results in Rabbits, Diabetes Technology& Therapeuticsl(2): 145-151 (1999). Erckens et al.; Raman Spectroscopy for Non-Invasive Characterization of Ocular Tissue: Potential for Detection of Biological Molecules, Abstract, J. of Raman Spectroscopy, 28(5):293-9 (May 1997). Schulze HG et al., Artificial Neural Network and Classical Least-S...
متن کاملDetermination of Insulin Concentration in Camel Milk Using Ultra Violet –Visible Absorption Spectroscopy
ABSTRACT: Camel milk contains sufficient levels of insulin and does not form the coagulum in the stomach preventing degradation of insulin and act as an effective vehicle to take the milk insulin unchanged to the intestine. In present study direct spectroscopy and standard addition method, well-known methods in analytical chemistry were used for determining insulin concentration in camel milk. ...
متن کاملDetermination of Kinetic Parameters for Catalytic Isomerization of Glucose to Fructose by immobilized Glucose Isomerase in an Aqueous-Ethanol Medium
Catalytic isomerization of glucose to fructose by immobilized glucose isomerase in an aqueous-ethanol medium was studied. Using Michaelis-Menten equation and Haldane relationship, the main kinetic parameters were determined: Kmf, Vmf, Kmr, Vmr. By comparing the aqueous-ethanol medium with an aqueous solution, the measured values of Kmr and Kmf were increased in the aqueous-ethanol solution, whi...
متن کاملDetermination of Kinetic Parameters for Catalytic Isomerization of Glucose to Fructose by immobilized Glucose Isomerase in an Aqueous-Ethanol Medium
Catalytic isomerization of glucose to fructose by immobilized glucose isomerase in an aqueous-ethanol medium was studied. Using Michaelis-Menten equation and Haldane relationship, the main kinetic parameters were determined: Kmf, Vmf, Kmr, Vmr. By comparing the aqueous-ethanol medium with an aqueous solution, the measured values of Kmr and Kmf were increased in the aqueous-ethanol solution, whi...
متن کاملDetection and Characterization of Human Teeth Caries Using 2D Correlation Raman Spectroscopy
Background: Carious lesions are formed by a complex process of chemical interaction between dental enamel and its environment. They can cause cavities and pain, and are expensive to fix. It is hard to characterize in vivo as a result of environment factors and remineralization by ions in the oral cavity. Objectives: The development of a technique that gives early diagnosis which is non-invasi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Applied spectroscopy
دوره 59 8 شماره
صفحات -
تاریخ انتشار 2005