Intraocular lens alignment from an en face optical coherence tomography image Purkinje-like method
نویسندگان
چکیده
Mengchan Sun Alberto de Castro Sergio Ortiz Pablo Perez-Merino Judith Birkenfeld Susana Marcos Instituto de Óptica “Daza de Valdés” Consejo Superior de Investigaciones Científicas Serrano 121 28006 Madrid, Spain E-mail: [email protected] Abstract. Measurement of intraocular lens (IOL) alignment implanted in patients in cataract surgery is important to understand their optical performance. We present a method to estimate tilt and decentration of IOLs based on optical coherence tomography (OCT) images. En face OCT images show Purkinje-like images that correspond to the specular reflections from the corneal and IOL surfaces. Unlike in standard Purkinjeimaging, the tomographic nature of OCT allows unequivocal association of the reflection with the corresponding surface. The locations of the Purkinje-like images are linear combinations of IOL tilt, IOL decentration, and eye rotation. The weighting coefficients depend on the individual anterior segment geometry, obtained from the same OCT datasets. The methodology was demonstrated on an artificial model eye with set amounts of lens tilt and decentration and five pseudophakic eyes. Measured tilt and decentration in the artificial eye differed by 3.7% and 0.9%, respectively, from nominal values. In patients, average IOL tilt and decentration from Purkinje were 3.30 4.68 deg and 0.16 0.16 mm, respectively, and differed on average by 0.5 deg and 0.09 mm, respectively, from direct measurements on distortion-corrected OCT images. Purkinje-based methodology from anterior segment en face OCT imaging provided, therefore, reliable measurements of IOL tilt and decentration. © The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI. [DOI: 10.1117/1.OE.53.6.061704]
منابع مشابه
Real time en face Fourier-domain optical coherence tomography with direct hardware frequency demodulation.
We demonstrate en face swept source optical coherence tomography (ss-OCT) without requiring a Fourier transformation step. The electronic optical coherence tomography (OCT) interference signal from a k-space linear Fourier domain mode-locked laser is mixed with an adjustable local oscillator, yielding the analytic reflectance signal from one image depth for each frequency sweep of the laser. Fu...
متن کاملOptical Coherence Tomography (OCT) of the Anterior Eye Segment
Description of Procedure or Service Optical coherence tomography (OCT) is a high resolution method of imaging the ocular structures. OCT for the anterior eye segment is being evaluated as a non-invasive diagnostic and screening tool for the detection of angle closure glaucoma, to assess corneal thickness and opacity, evaluate pre-surgical and postsurgical anterior chamber anatomy, calculate int...
متن کاملIntraocular lens alignment from purkinje and Scheimpflug imaging.
The improved designs of intraocular lenses (IOLs) implanted during cataract surgery demand understanding of the possible effects of lens misalignment on optical performance. In this review, we describe the implementation, set-up and validation of two methods to measure in vivo tilt and decentration of IOLs, one based on Purkinje imaging and the other on Scheimpflug imaging. The Purkinje system ...
متن کاملEvaluation of scleral-fixated intraocular lens position anomalies by anterior segment optical coherence tomography.
BACKGROUND/AIM To measure the positions of scleral-fixated intraocular lenses (IOLs) using anterior segment optical coherence tomography (OCT) and to evaluate the influence of the position of scleral-fixated IOL on the quality of vision. MATERIALS AND METHODS Twenty-six eyes of 26 patients were included in the study. The average age was 57.5 (11 to 78) years. Anterior segment images were take...
متن کاملSwept Source Optical Coherence Tomography Angiography for Contact Lens-Related Corneal Vascularization
Purpose. To describe a novel technique of adapting a swept-source optical coherence tomography angiography (OCTA) to image corneal vascularization. Methods. In this pilot cross-sectional study, we obtained 3 × 3 mm scans, where 100,000 A-scans are acquired per second with optical axial resolution of 8 μm and lateral resolution of 20 μm. This was performed with manual "XYZ" focus without the ant...
متن کامل