Use of a multileaf collimator to increase the field width achievable with a dynamic wedge
نویسندگان
چکیده
A method is proposed for generating dynamic wedges spanning the entire field width, defined as the collimator opening in the wedged direction, without changes to existing hardware. The technique approximates the fluence pattern of a dynamic wedge by sequentially closing the leaves of a 120-leaf multileaf collimator (MLC). Closure times for the individual leaves were derived by extending the segmented treatment table of the dynamic wedge provided by the manufacturer of the linear accelerator. Using film dosimetry, beam properties of MLC wedges were compared to those of conventional dynamic and mechanical wedges. Profiles and isodose lines of the MLC wedge were almost identical to those of the dynamic wedge, and differed only modestly from the mechanical counterparts. Dose inhomogeneity due to the individually closing leaves was not significant. The high-dose region at the junction between opposing MLC leaves, unavoidable when the field length (i.e., the opening of the collimator in the nonwedged direction) exceeds the maximum leaf extension of 15 cm, was feathered by moving leaf pairs after their closure for the remainder of the irradiation time. Combining the MLC wedge with a regular dynamic wedge to reduce the line of high dose under the leaf junction is under consideration.
منابع مشابه
Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D
Background:Â Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accur...
متن کاملGeneration of wedge‐shaped dose distributions through dynamic multileaf collimator dose delivery
A new method of generating wedge-shaped dose distributions through dynamic multileaf collimator dose delivery rather than computer-controlled jaw motion is presented. The method starts with the calculation of a wedge-shaped beam profile for the desired wedge angle. The resultant wedge beam profile is then passed to the intensity-modulated radiation therapy (IMRT) leaf sequence generation algori...
متن کاملMonte Carlo Study of Unflattened Photon Beams Shaped by Multileaf Collimator
Introduction: This study investigates basic dosimetric properties of unflattened 6 MV photon beam shaped by multileaf collimator and compares them with those of flattened beams.Materials and Methods: Monte Carlo simulation model using BEAM code was developed for a 6MV photon beam based on Varian Clinic 600 unique performance linac operated with and without a flattening filter in beam line. Dosi...
متن کاملThe Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions
Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans. Materials and Method...
متن کاملDesign and Fabrication of the Mechanical Part of a Prototype Multileaf Collimator System
Introduction: Multileaf collimator (MLC) is state of the art among the radiation field shaping systems for conformal radiotherapy techniques. Some advantages of this system are as follow: reduction of treatment time, higher precision in conforming the radiation field to the projection of the target volume and lowering the chance of errors made by technologists. The purpose of this research was ...