Evaluation of pleural and ascitic fluid analysis on the Sysmex XE-5000 hematology analyzer

نویسندگان

  • Samuel Ricardo Comar
  • Angela Maria de Souza
  • Gisele Maria B. Singer
چکیده

Introduction: Body fluid (BF) analysis is critical to the diagnosis and monitoring of several pathological conditions. The limitations of manual cell counts have led to greater interest in the development of automated BF analysis. Objective: To evaluate the analytical performance of the Sysmex XE-5000 hematology analyzer in the analysis of pleural and ascitic fluids in the laboratory routine of a large university hospital. Methods: A total of 56 samples (35 ascitic and 21 pleural fluids) were analyzed by manual optical microscopy (OM) and XE-5000. Analytical performance includes linearity, carryover, functional sensitivity and comparison of patient samples. Results: Performance studies showed linearity up to 25,825 WBC-BF/μl (r = 0.999), WBC-BF showed carryover of 0.18%, and the lower limit of quantitation was set at 22 WBC-BF/μl. Good correlations between the methods were observed just for total cell (TC-BF) and white blood cell (WBC-BF) counts in pleural and ascitic fluids. The high-fluorescence cell count (HF-BF) showed poor correlation but high positive predictive value (PPV) for both fluids (94.74% for pleural and 96.97% for ascitic fluid). Conclusion: XE-5000 provides accurate and precise count for TC-BF, WBC-BF and polymorphonuclear cells (PMN-BF) in pleural and ascitic fluids in medical decision levels, but the morphological differentiation should continue to be held by OM. Histogram and scattergram displayed on XE-5000 must always be analyzed to assess if there is any interference or flag. The HF-BF parameter is a potential tool for screening.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance characteristics of the Sysmex XT-2000i hematology analyzer.

The XT-2000i is a new hematology analyzer combining aperture impedance and optical technologies. It is capable of generating a 30-parameter hematologic assay on peripheral blood at a throughput of 80 samples per hour. We have evaluated the performance characteristics of the XT-2000i analyzer and compared them with those of the state-of-the-art XE-2100 analyzer and the manual differential count....

متن کامل

Evaluation of the Diagnostic Performance of theSysmex XT-2000i Automated Hematology Analyzerin the Detection of Immature Granulocyte

Immature Granulocytes (IG) are normally not present in peripheral blood. Their presence is an important indicator of enhanced bone marrow activation. The present study evaluated the diagnostic performance of the Sysmex XT-2000i automated hematology analyzer in the detection of myelocytes and/or metamyelocytes in patients affected by non-haematological disease, in comparison with manual microsco...

متن کامل

Performance evaluation of the immature granulocyte parameter on the Sysmex XE-2100 automated hematology analyzer.

Recently, St. Anthony's Medical Center Laboratory (SAMC) in St. Louis, Missouri, installed a new automated hematology analyzer, the Sysmex XE-2100. The XE-2100 has an additional differential parameter, the immature granulocyte (IG) parameter. Prior to implementation, the IG parameter was evaluated by comparison to the manual immature granulocyte count. Immature granulocytes counted manually, as...

متن کامل

Overview of Automated Hematology Analyzer XE-2100

The XE-2100 is a new automated hematology analyzer developed by Sysmex Corporation. With consideration of medical analyzer requirements for the 21st century, this unit was developed to satisfy the following concepts and to cope with diversified laboratory needs. 1) Electronic network communication 2) Presentation of high level value-added information 3) Contribution to test cost control 4) Purs...

متن کامل

Preliminary Data on the Feasibility of Bone Marrow Screening on the Sysmex XE-2100 Automated Hematology Analyzer

In the present study we investigated the potential of the Sysmex XE-2100 automated hematology analyzer as a screening device for bone marrow (BM) aspirates. We evaluated BM cellularity, white blood cell (WBC) differential count, immature granulocyte (IG) count, nucleated red blood cell (NRBC) count, granulocyte/erythroid (G/E) and myeloid/erythroid (M/E) ratio, and potential pathologyrelated sc...

متن کامل