Measurement of respiratory impedance by forced oscillation: comparison of the standard and head generator methods.

نویسندگان

  • Y Iwatsubo
  • H Lorino
  • C Hubert
  • C Duvivier
  • R Peslin
  • Q T Pham
  • T Moreau
  • J J Hosselet
  • P Brochard
چکیده

Physiological and clinical studies have shown that the standard method of measuring respiratory impedance by forced oscillation leads to less efficient control of the upper airway shunt effect than the head generator method. To test the effects of these two techniques in epidemiological studies, we compared, in a sample of 73 French agricultural workers, the values obtained with each method for five forced oscillation parameters: resistance, frequency dependence of resistance, inertance, compliance and resonant frequency. For these comparisons, subjects were classified according to four respiratory status factors: smoking status, cough, expectoration and airway obstruction assessed from the maximum expiratory flow volume curve. Logistic regression models using the set of four forced oscillation parameters (excluding resonant frequency, which is derived from compliance and inertance) were then used to analyse the ability of each method to classify the subjects in each group. Significant differences between the two methods were observed for the mean values obtained for all five parameters. However, when each parameter was considered separately, the correlations between the values for each method were significant. Each method possessed the necessary ability to separate subjects into our group classification, but the significant relationships were not always found for the same parameters. Finally, logistic regression models showed that the two methods led to almost the same classification of the subjects. According to our results, the standard method of applying forced oscillation at the mouth seems an acceptable device for measuring respiratory impedance for epidemiological purposes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Input respiratory impedance to estimate airway hyperreactivity in children: standard method versus head generator.

We previously found that a significant underestimation of respiratory mechanical impedance (Zrs) at high frequency may result from the upper airway artefact in children, when pressure is directly varied at the mouth. To determine the importance of this artefact in estimating lung response to bronchomotor agents with the forced oscillation technique, input respiratory mechanical impedance was me...

متن کامل

Influence of upper airway shunt on total respiratory impedance in infants.

When input impedance is determined by means of the forced oscillation technique, part of the oscillatory flow measured at the mouth is lost in the motion of the upper airway wall acting as a shunt. This is avoided by applying the oscillations around the subject's head (head generator) rather than at the mouth (conventional technique). In seven wheezing infants, we compared both techniques to es...

متن کامل

Servocontrolled generator to measure respiratory impedance from 0.25 to 26 Hz in ventilated patients at different PEEP levels.

Assessing respiratory impedance (Zrs) in ventilated patients over a wide frequency band, ranging from breathing rates to typical forced oscillation frequencies, during end-expiratory pauses at different positive end-expiratory pressures (PEEP) is of potential interest to assess a patient's respiratory mechanics. Zrs measurements under these conditions are not possible with the present variants ...

متن کامل

Assessment of bronchial reactivity by forced oscillation admittance avoids the upper airway artefact.

The forced oscillation technique (FOT) allows easy assessment of bronchial reactivity. The use of a standard FOT generator (SG) results in changes in respiratory system resistance (delta Rrs,SG) which are affected by an artefact caused by the extrathoracic upper airway (EUA). The aim was to improve the FOT assessment of bronchial reactivity with the SG by computing the change in FOT admittance ...

متن کامل

Forced oscillation techniques for measuring mechanical properties of the respiratory system.

Airway function is usually assessed by tests of forced expiration or, in more specialised laboratories, by measuring airways resistance by whole body plethysmography. Plethysmography depends on the patients' cooperation and requires expensive and bulky apparatus. An alternative non-invasive technique for measuring airflow resistance, using forced oscillation, was introduced by DuBois and collea...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 7 5  شماره 

صفحات  -

تاریخ انتشار 1994