Monte Carlo‐based beam quality and phantom scatter corrections for solid‐state detectors in 60Co and 192Ir brachytherapy dosimetry

نویسندگان

  • Mishra Subhalaxmi
  • T. Palani Selvam
چکیده

Beam quality correction, kQQ0(r), for solid-state detectors diamond, LiF, Li2B4O7, Al2O3, and plastic scintillator are calculated as a function of distance, r, along the transverse axis of the 60Co and 192Ir brachytherapy sources using the Monte Carlo- based EGSnrc code system. This study also includes calculation of detector-specific phantom scatter correction, kphan(r), for solid phantoms such as PMMA, polysty- rene, solid water, virtual water, plastic water, RW1, RW3, A150, and WE210. For 60Co source, kQQ0(r) is about unity and distance-independent for diamond, plastic scintillator, Li2B4O7 and LiF detectors. For this source, kQQ0(r) decreases gradually with r for Al2O3 detector (about 6% smaller than unity at 15 cm). For 192Ir source, kQQ0(r) is about unity and distance-independent for Li2B4O7 detector (overall variation is about 1% in the distance range of 1-15 cm). For this source, kQQ0(r) increases with r for diamond and plastic scintillator (about 6% and 8% larger than unity at 15 cm, respectively). Whereas kQQ0(r) decreases with r gradually for LiF (about 4% smaller than unity at 15 cm) and steeply for Al2O3 (about 25% smaller than unity at 15 cm). For 60Co source, solid water, virtual water, RW1, RW3, and WE210 phantoms are water-equivalent for all the investigated solid-state detectors. Whereas polystyrene and plastic water phantoms are water-equivalent for diamond, plastic scintillator, Li2B4O7 and LiF detectors, but show distance-dependent kphan(r) values for Al2O3 detector. PMMA phantom is water-equivalent at all distances for Al2O3 detector, but shows distance-dependent kphan(r) values for remaining detec- tors. A150 phantom shows distance-dependent kphan(r) values for all the investigated detector materials. For 192Ir source, solid water, virtual water, RW3, and WE210 phantoms are water-equivalent for diamond, plastic scintillator, Li2B4O7 and LiF detectors, but show distance-dependent kphan(r) values for Al2O3 detector. All other phantoms show distance-dependent kphan(r) values for all the detector materials. 

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo Calculation of Beam Quality and Phantom Scatter Corrections for Lithium Formate Electron Paramagnetic Resonance Dosimeter for High-energy Brachytherapy Dosimetry

PURPOSE To investigate beam quality correction, KQQ0 (r) and phantom scatter correction, Kphan (r) for lithium formate dosimeter as a function of distance r along the transverse axis of the high-energy brachytherapy sources 60Co, 137Cs, 192Ir and 169Yb using the Monte Carlo-based EGSnrc code system. MATERIALS AND METHODS The brachytherapy sources investigated in this study are BEBIG High Dose...

متن کامل

Monte Carlo calculation of beam quality correction for solid‐state detectors and phantom scatter correction at 137Cs energy

Beam quality correction kQQo (r), which reflects the absorbed energy dependence of the detector, is calculated for solid state detector materials diamond, LiF, Li2B4O7 and Al2O3 for the 137Cs RTR brachytherapy source using the Monte Carlo-based EGSnrc code system. The study also includes calculation of detector-specific phantom scatter corrections kphant (r) for solid phantoms such as PMMA, pol...

متن کامل

Determination of TG-43 Dosimetric Parameters for Photon Emitting Brachytherapy Sources

Objective: Brachytherapy sources are widely used for the treatment of cancer. The report of Task Group No. 43 (TG-43) of American Association of Physicists in Medicine is known as the most common method for the determination of dosimetric parameters for brachytherapy sources. The aim of this study is to obtain TG-43 dosimetric parameters for 60Co, 137Cs, 192Ir and 103Pd brachyt...

متن کامل

Monte Carlo‐based investigation of absorbed‐dose energy dependence of radiochromic films in high energy brachytherapy dosimetry

Relative absorbed dose energy response correction, R, for various radiochromic films in water phantom is calculated by the use of the Monte Carlo-based EGSnrc code system for high energy brachytherapy sources 60Co, 137Cs, 192Ir and 169Yb. The corrections are calculated along the transverse axis of the sources (1-15 cm). The radiochromic films investigated are EBT, EBT2 (lot 020609 and lot 03110...

متن کامل

Monte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost

Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phan...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 15  شماره 

صفحات  -

تاریخ انتشار 2014