To evaluate the capability of magnetic resonance (MR) in imaging normal acoustic nerves, 12 volunteers without signs or symptoms of intracranial disease were examined using a 0.6 T superconductive system. Several spin-echo (SE) pulse sequences were tested to identify the optimal sequence for demonstration of the acoustic nerve bundle. Repetition times (TRs) varied from 300 to 2000 msec and echo times (TEs) from 30 to 120 msec. A single-slice technique was used with 5 and 8 mm sections, one or two data acquisitions per projection, and axial and coronal imaging. The normal acoustic nerves were demonstrated readily by MR in axial and/or coronal sections. The distal parts of the nerves and tumors were imaged best with SE 1500/60. The medial extremities of the seventh and eighth nerves tended to be obscured in this sequence by brightening the cerebrospinal fluid signal adjacent to the brainstem, but they were demonstrated clearly with 500 or 800 msec TR and 30 msec TE. Five patients were studied who had hearing loss and evidence of retrocochlear disease. In four patients, MR imaging demonstrated five acoustic nerve tumors ranging in size from purely intracanalicular to a 12 mm cisternal component. In the fifth case, no tumor was identified by MR imaging or gas computed tomographic (CT) cisternography. Contrast-enhanced CT using a Siemens Somatom DR 3 or GE CT/T 8800 scanner failed to provide convincing evidence of tumor in any case, while gas CT cisternography was positive in all five tumors. All five acoustic neuromas were identified readily using the SE sequences that proved optimal for demonstration of normal nerves. This experience revealed that MR imaging can demonstrate the eighth nerve complex well and reliably. Single-slice (5 or 8 mm) technique is adequate, but multislice without tissue gaps (used recently) is more efficient. Small, even intracanalicular, acoustic neuromas are imaged effectively, indicating that the method is capable of superseding contrast CT cisternography, particularly with improving technology.
منابع مشابه
Surface-coil magnetic resonance imaging of the internal auditory canal.
Computed tomography is effective for detecting acoustic neuromas, but not for resolving individual nerves in the internal auditory canal. Surface-coil magnetic resonance (MR) images of the internal auditory canal were obtained using a 1.5 T superconducting magnet, a 13.5-cm-diameter surface coil, 3- and 5-mm-thick slices, and partial-saturation pulse sequences. Cranial nerves VII and VIII (thre...
متن کاملAcoustic Neurinomas
Acoustic neuromas (AN) are schwann cell-derived tumors that commonly arise from the vestibular portion of the eighth cranial nerve also known as vestibular schwannoma(VS) causes unilateral hearing loss, tinnitus, vertigo and unsteadiness. In many cases, the tumor size may remain unchanged for many years following diagnosis, which is typically made by MRI. In the majority of cases the tumor is s...
متن کاملThe role of magnetic resonance imaging in the identification of suspected acoustic neuroma: a systematic review of clinical and cost effectiveness and natural history.
OBJECTIVE(S) To evaluate the clinical effectiveness and cost-effectiveness of a range of diagnostic strategies for investigating patients with unilateral hearing loss and/or tinnitus, with a view to confirming or eliminating a diagnosis of acoustic neuroma, and to describe the natural history of acoustic neuroma. DATA SOURCES Major electronic databases were searched from January 1980 to Augus...
متن کاملThe sensitivity of auditory brainstem response testing for the diagnosis of acoustic neuromas.
OBJECTIVES To determine the sensitivity of auditory brainstem response (ABR) testing for detecting acoustic neuromas and to determine whether the test is less sensitive for detecting small tumors. DESIGN Retrospective review of the medical charts of 58 patients with acoustic neuroma who had all of the data necessary for inclusion in the study. SETTING University-affiliated referral practice...
متن کاملMR imaging of cerebellopontine angle and internal auditory canal lesions at 1.5 T.
The high-field, thin-section (3-5 mm) MR imaging characteristics of 49 cerebellopontine angle and internal auditory canal lesions were reviewed. The diverse abnormalities include 20 acoustic neurinomas, eight neurinomas of other cranial nerves (six involving the fifth cranial nerve and two involving cranial nerves IX-XI), seven meningiomas, five subdural fluid collections, four brainstem tumors...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- AJR. American journal of roentgenology
دوره 144 5 شماره
صفحات -
تاریخ انتشار 1985