Construction and Accuracy Assessment of Patient-Specific Biocompatible Drill Template for Cervical Anterior Transpedicular Screw (ATPS) Insertion: An In Vitro Study
نویسندگان
چکیده
BACKGROUND With the properties of three-column fixation and anterior-approach-only procedure, anterior transpedicular screw (ATPS) is ideal for severe multilevel traumatic cervical instabilities. However, the accurate insertion of ATPS remains challenging. Here we constructed a patient-specific biocompatible drill template and evaluated its accuracy in assisting ATPS insertion. METHODS After ethical approval, 24 formalin-preserved cervical vertebrae (C2-C7) were CT scanned. 3D reconstruction models of cervical vertebra were obtained with 2-mm-diameter virtual pin tracts at the central pedicles. The 3D models were used for rapid prototyping (RP) printing. A 2-mm-diameter Kirschner wire was then inserted into the pin tract of the RP model before polymethylmethacrylate was used to construct the patient-specific biocompatible drill template. After removal of the anterior soft tissue, a 2-mm-diameter Kirschner wire was inserted into the cervical pedicle with the assistance of drill template. Cadaveric cervical spines with pin tracts were subsequently scanned using the same CT scanner. A 3D reconstruction was performed of the scanned spines to get 3D models of the vertebrae containing the actual pin tracts. The deviations were calculated between 3D models with virtual and actual pin tracts at the middle point of the cervical pedicle. 3D models of 3.5 mm-diameter screws were used in simulated insertion to grade the screw positions. FINDINGS The patient-specific biocompatible drill template was constructed to assist ATPS insertion successfully. There were no significant differences between medial/lateral deviations (P = 0.797) or between superior/inferior deviations (P = 0.741). The absolute deviation values were 0.82±0.75 mm and 1.10±0.96 mm in axial and sagittal planes, respectively. In the simulated insertion, the screws in non-critical position were 44/48 (91.7%). CONCLUSIONS The patient-specific drill template is biocompatible, easy-to-apply and accurate in assisting ATPS insertion. Its clinical applications should be further researched.
منابع مشابه
A Novel Patient-Specific Drill Guide Template for Pedicle Screw Insertion into the Subaxial Cervical Spine Utilizing Stereolithographic Modelling: An In Vitro Study
STUDY DESIGN Cadaveric study. PURPOSE The purpose of this study was to assess the accuracy and feasibility of cervical pedicle screw (CPS) insertion into the subaxial cervical spine placed using a patient-specific drill guide template constructed from a stereolithographic model. OVERVIEW OF LITERATURE CPS fixation is an invaluable tool for posterior cervical fixation because of its biomecha...
متن کاملAccuracy of fluoroscopy versus computer-assisted navigation for the placement of anterior cervical pedicle screws.
STUDY DESIGN Randomized laboratory cadaver study. OBJECTIVE The objective of this study was to determine the accuracy of anterior transpedicular screw placement in the cervical spine using conventional fluoroscopy versus computer-assisted navigation. SUMMARY OF BACKGROUND DATA Traditionally, global cervical instability has required anterior and posterior fixation due to the superior biomech...
متن کاملCT Morphometric Analysis to Determine the Anatomical Basis for the Use of Transpedicular Screws during Reconstruction and Fixations of Anterior Cervical Vertebrae
BACKGROUND Accurate placement of pedicle screw during Anterior Transpedicular Screw fixation (ATPS) in cervical spine depends on accurate anatomical knowledge of the vertebrae. However, little is known of the morphometric characteristics of cervical vertebrae in Chinese population. METHODS Three-dimensional reconstructions of CT images were performed for 80 cases. The anatomic data and screw ...
متن کاملA novel patient-specific navigational template for cervical pedicle screw placement.
STUDY DESIGN.: Prospective trial. OBJECTIVE.: To develop and validate a novel, patient-specific navigational template for cervical pedicle placement. SUMMARY OF BACKGROUND DATA.: Owing to the narrow bony anatomy and the proximity to the vertebral artery and the spinal cord, cervical instrumentation procedures demand the need for a precise technique for screw placement. PATIENT.: Specific drill ...
متن کاملAccuracy Assessment of Freehand Pedicular Screw Placement
Background: The purpose of this prospective study was to determine the accuracy of pedicular screw insertion withoutthe use of fluoroscopy.Methods: This study was conducted on patients with spinal diseases in need of pedicular screw fixation and fusion.The included patients suffered from such conditions as vertebral fracture, spinal stenosis, kyphosis, tumor, and pelvicf...