Biophysical characterization of a relativistic proton beam for image-guided radiosurgery
نویسندگان
چکیده
We measured the physical and radiobiological characteristics of 1 GeV protons for possible applications in stereotactic radiosurgery (image-guided plateau-proton radiosurgery). A proton beam was accelerated at 1 GeV at the Brookhaven National Laboratory (Upton, NY) and a target in polymethyl methacrylate (PMMA) was used. Clonogenic survival was measured after exposures to 1-10 Gy in three mammalian cell lines. Measurements and simulations demonstrate that the lateral scattering of the beam is very small. The lateral dose profile was measured with or without the 20-cm plastic target, showing no significant differences up to 2 cm from the axis A large number of secondary swift protons are produced in the target and this leads to an increase of approximately 40% in the measured dose on the beam axis at 20 cm depth. The relative biological effectiveness at 10% survival level ranged between 1.0 and 1.2 on the beam axis, and was slightly higher off-axis. The very low lateral scattering of relativistic protons and the possibility of using online proton radiography during the treatment make them attractive for image-guided plateau (non-Bragg peak) stereotactic radiosurgery.
منابع مشابه
The Mobius AIRO mobile CT for image‐guided proton therapy: Characterization & commissioning
PURPOSE The purpose of this study was to characterize the Mobius AIRO Mobile CT System for localization and image-guided proton therapy. This is the first known application of the AIRO for proton therapy. METHODS Five CT images of a Catphan® 504 phantom were acquired on the AIRO Mobile CT System, Varian EDGE radiosurgery system cone beam CT (CBCT), Philips Brilliance Big Bore 16 slice CT simu...
متن کاملThe Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography
Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...
متن کاملAccurate Proton Beam Localization
The goal of this work was to further improve an experimental proton radiosurgery system at Loma Linda University Medical Center to reach sub-millimeter accuracy before proton radiosurgery with narrow beams can be used in a clinical trial. Radiosurgery precisely targets a specific anatomical region with high doses of radiation. We have developed a program that provides correcting translational o...
متن کاملVerification procedure for isocentric alignment of proton beams
We present a technique--based on the Lutz, Winston, and Maleki test used in stereotactic linear accelerator radiosurgery--for verifying whether proton beams are being delivered within the required spatial coincidence with the gantry mechanical isocenter. Our procedure uses a proton beam that is collimated by a circular aperture at its central axis and is then intercepted by a small steel sphere...
متن کاملClinical trials and radiation oncology technologies.
At the very start of their clinical training, radiation oncologists tend to embrace technology. The specialty draws a preponderance of technophiles to it, and those few who are not at first passionate about technology quickly become so. This has been especially true over the past 2 decades with the introduction of innovative therapies such as three-dimensional conformal radiation therapy (3DCRT...