Preservation of distal coronary perfusion during prolonged balloon inflation with an autoperfusion angioplasty catheter.
نویسندگان
چکیده
A newly designed balloon coronary angioplasty catheter that allows passive antegrade blood flow during balloon inflation (autoperfusion catheter) was compared with a standard balloon coronary angioplasty catheter. In a randomized sequence, inflations were performed for 3 min in the left circumflex coronary artery of 12 dogs with the standard catheter followed by the autoperfusion catheter or vice versa. During inflation with the standard catheter, the ST segment of standard limb lead II increased from -0.02 +/- 0.03 mV to 0.39 +/- 0.08 mV (p less than .001), whereas during inflation with the autoperfusion catheter the ST segment did not change (-0.03 +/- 0.03 vs -0.01 +/- 0.04 mV; p = NS). Regional myocardial blood flow measured by the radioactive microsphere technique in the posterior subepicardium and subendocardium was 0.12 +/- 0.03 and 0.08 +/- 0.03 ml/min/g, respectively, with the standard catheter as compared with 0.57 +/- 0.08 and 0.61 +/- 0.14 ml/min/g with the autoperfusion catheter (both p less than .01 compared with the standard catheter). Thus, unlike the standard catheter, the autoperfusion catheter allows for inflations up to 3 min in duration without producing deleterious changes in the ST segment or severe reductions in regional myocardial blood flow.
منابع مشابه
Myocardial protection during coronary angioplasty with an autoperfusion balloon catheter in humans.
An autoperfusion balloon catheter was developed to allow passive myocardial perfusion during inflation through a central lumen and multiple side holes in the shaft proximal and distal to the balloon. We report its safety and efficacy in 11 patients undergoing elective angioplasty to a single coronary lesion. Each lesion was dilated three times with the autoperfusion inflation bracketed between ...
متن کاملExtent of Electrocardiographic Ischemic Changes and Duration of Inflation With Standard and Autoperfusion Angioplasty SRI PBC SR
An autoperfusion balloon catheter was developed to allow passive myocardial perfusion during inflation through a central lumen and multiple side holes in the shaft proximal and distal to the balloon. We report its safety and efficacy in 11 patients undergoing elective angioplasty to a single coronary lesion. Each lesion was dilated three times with the autoperfusion inflation bracketed between ...
متن کاملExtent of Electrocardiographic Ischemic Changes and Duration of Inflation With Standard and Autoperfusion Angioplasty SRI PBC SR 2 ST
An autoperfusion balloon catheter was developed to allow passive myocardial perfusion during inflation through a central lumen and multiple side holes in the shaft proximal and distal to the balloon. We report its safety and efficacy in 11 patients undergoing elective angioplasty to a single coronary lesion. Each lesion was dilated three times with the autoperfusion inflation bracketed between ...
متن کاملCoronary artery dissection caused by exit of the guidewire through the distal perfusion sidehole of an auto-perfusion angioplasty balloon catheter.
A previously unrecognized cause of coronary artery dissection is reported. A 67-year-old woman underwent angioplasty of the right coronary artery using an autoperfusion balloon catheter. Dissection occurred because the balloon catheter was advanced while the guidewire exited from one of the distal perfusion sideholes.
متن کاملDetection of coronary collateral flow by a Doppler-tipped guide wire during coronary angioplasty.
Coronary collaterals may not be apparent during routine coronary angiography or during coronary angioplasty as a result of a variety of physiologic and anatomic factors.‘-” The dynamic nature of the collateral circulation has been demonstrated by newly appearing angiographic collaterals during contralateral vessel occlusion.3 An elevated coronary occlusion wedge pressure during angioplasty ball...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Circulation
دوره 75 6 شماره
صفحات -
تاریخ انتشار 1987