An evaluation of three methods for determining colloid osmotic pressure.

نویسندگان

  • M M Blackwell
  • J Riley
  • M McCall
  • J Ecklund
  • R Southworth
چکیده

Plasma colloid osmotic pressure (COP) is an important determinant in edema formation. Three methods for assessing the COP were evaluated. Direct measurement of COP using the 4420 Wescor Colloid Osmometer was compared to the estimation of COP from both serum total protein and total serum solids (TSS) determinations. Blood samples from twenty adult patients (mean age = 64 years) undergoing cardiopulmonary bypass surgery were collected for COP assessment. Sample collection was performed prior to heparinization/hemodilution, during hypothermic bypass and at the conclusion of bypass following protamine administration. The results obtained from each method were analyzed by a two-way analysis of variance. The Bonferroni technique was used for comparison of sample means when the difference was significant (p less than 0.05). Correlations were reported by linear regression analysis. A statistically significant difference (p less than 0.01) was found between the three methods. A regression equation for the estimation of COP from total serum solids is offered: COP = (3.02 * TSS) + 0.65. Prospective clinical testing between the direct COP measurement and the estimation of COP from TSS using the equation (n = 38) revealed a significant correlation (R2 = .932) and no significant difference between the two (p greater than 0.05).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Determination of the Colloidal Osmotic Pressure in Blood Serum and Similar Fluids

The importance of the colloidal osmotic pressure in biological fluids is becoming increasingly recognized, but difficulties in the older methods as yet prevent their general use in inquiries for which needed information can only be obtained from colloidal osmotic pressure data. The membrane bag arrangement used by Starling (1896) is capable of yielding admirable results (see, e.g., Adair, 1925;...

متن کامل

Stability of Colloid Osmotic Pressure of Serum Protein By

The work of earlier investigators, Govaerts (1), Barath (2), Cope (3), Peters et al. (4) and Kylin (5) has given the impression that the colloid osmotic pressure is not only lowered with decrease in serum protein but its specific pressure (pressure per unit of protein) is also diminished. The lowering of specific pressure has usually been related to a greater decrease in the albumin fraction of...

متن کامل

The Validity of Determinations of the Colloid Osmotic Pressure of Serum

Measurement of the colloid osmotic pressure of the blood, or argument depending thereon, has been involved in much recent work on the origin of tissue fluid from the blood through capillary walls and in work on changes taking place in the blood due to various causes. It may be sufficient to mention Krogh (1929), Landis (1925-26, 1927, 1930, a, 1930, b), Loewen, Field, and Drinker (1931), Meyer ...

متن کامل

Transcapillary fluid flux and inflammatory response during neonatal therapeutic hypothermia: an open, longitudinal, observational study

BACKGROUND Therapeutic hypothermia is neuroprotective in asphyxiated neonates by counteracting mechanisms contributing to brain injury. Although an initial increased permeability is part of an inflammatory reaction and thereby a natural healing process, an excessive endothelial permeability with edema formation may result in impaired hemodynamics. Reduced permeability may, however, benefit heal...

متن کامل

Albumin/globulin rations, colloid osmotic pressures and Lowry protein microanalysis in rat plasma.

1. We have correlated measurements of colloid osmotic pressure with protein concentrations and albumin/globulin ratios. Plasma from normal rats of different genetic strains was used. 2. The Landis and Pappenheimer equation does not accurately predict colloid osmotic pressure properties of the normal rat plasma used, which had an albumin/globulin ratio of 0.74. Over the range 2-2-12-2 g of prote...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 26 1  شماره 

صفحات  -

تاریخ انتشار 1994