Effect of Different Distal Fixation Augmentation Methods on the Pullout Strength of Fassier-Duval Telescoping Rods.
نویسندگان
چکیده
Antegrade telescoping rods have been introduced for use in pediatric patients with osteogenesis imperfecta (OI) to decrease the incidence of long-bone fractures and to correct and prevent deformities. Recent studies have documented failures of telescoping intramedullary rods due to inadequate distal fixation. The purpose of this study was to evaluate the pullout strength of distal fixation of the telescoping rod with and without synthetic calcium phosphate or polymethylmethacrylate (PMMA) augmentation. Four sets of 6 telescoping distal fixation rods were fixed according to standard insertion technique into an open-cell rigid-foam synthetic bone block simulating OI bone. The groups tested were as follows: control (no augmentation), 0.75 mL of PMMA-augmented, 0.75 mL of PMMA-rescued (stripped distal fixation, then resecured after PMMA augmentation), and 0.75 mL of bioabsorbable-calcium phosphate (CP)-augmented. All rods were tested to failure. The peak load was recorded. Average pullout strengths were as follows: control, 20±6.6 N; PMMA, 125±16.8 N; PMMA-rescued, 137±11.9 N; bioabsorbable-CP, 81±10.3 N. All augmented groups had significantly higher pullout strength compared with the control (P<.001). The PMMA and PMMA-rescued groups failed at the PMMA/bone interface, whereas the bioabsorbable-CP group failed at the cement/rod interface. All augmented constructs improved pullout strength by at least 400% compared with the control. Bioabsorbable cement may be less detrimental to the physis if pullout still occurs despite augmentation due to its mode of failure. This study provides biomechanical evidence to support the further in vivo investigation of either PMMA or bioabsorbable cement augmentation to improve pullout strength of distal telescoping rod fixation.
منابع مشابه
Advantages and Complications Following Fassier-Duval Intramedullary Rodding in Children. Pilot Study.
BACKGROUND The report presents the authors' experience with Fassier-Duval rods in children and an analysis of advantages and complications. MATERIAL AND METHODS Over a period of 15 months, the authors operated on 10 children and 18 rod implantation procedures were performed. In five cases, the implantation was a primary procedure, while in the remaining 13 cases Fassier-Duval rods were used t...
متن کاملCalcium phosphate cement augmentation of cancellous bone screws can compensate for the absence of cortical fixation.
An obvious means to improve the fixation of a cancellous bone screw is to augment the surrounding bone with cement. Previous studies have shown that bone augmentation with Calcium Phosphate (CaP) cement significantly improves screw fixation. Nevertheless, quantitative data about the optimal distribution of CaP cement is not available. The present study aims to show the effect of cement distribu...
متن کاملUse of minimally invasive (percutaneous) Fassier-Duval telescopic rod on an 8 year old patient with Lobstein disease.
Lobstein disease is a severe disease characterized primarily by bone fragility which causes numerous fractures, especially in long bones. Multiple fracture of a bone will cause in time a deformity with consecutive functional impotence. The treatment for Lobstein disease in children must take into account both increasing bone strength and fracture prevention and treatment of possible fractures. ...
متن کاملA Biomechanical Comparison of Expansive Pedicle Screws for Severe Osteoporosis: The Effects of Screw Design and Cement Augmentation
Expansive pedicle screws significantly improve fixation strength in osteoporotic spines. However, the previous literature does not adequately address the effects of the number of lengthwise slits and the extent of screw expansion on the strength of the bone/screw interface when expansive screws are used with or without cement augmentation. Herein, four designs for expansive pedicle screws with ...
متن کاملComparison of the Pullout Strength of Different Pedicle Screw Designs and Augmentation Techniques in an Osteoporotic Bone Model
Study Design Mechanical study. Purpose To compare the pullout strength of different screw designs and augmentation techniques in an osteoporotic bone model. Overview of Literature Adequate bone screw pullout strength is a common problem among osteoporotic patients. Various screw designs and augmentation techniques have been developed to improve the biomechanical characteristics of the bone-...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Orthopedics
دوره 39 2 شماره
صفحات -