Towards comprehensive characterization of Cs-137 Seeds using PRESAGE® dosimetry with optical tomography

نویسندگان

  • J Adamson
  • Y Yang
  • L Rankine
  • J Newton
  • J Adamovics
  • O Craciunescu
  • M Oldham
چکیده

We describe a method to directly measure the radial dose and anisotropy functions of brachytherapy sources using polyurethane based dosimeters read out with optical CT. We measured the radial dose and anisotropy functions for a Cs-137 source using a PRESAGE® dosimeter (9.5cm diameter, 9.2cm height) with a 0.35cm channel drilled for source placement. The dosimeter was immersed in water and irradiated to 5.3Gy at 1cm. Pre- and post-irradiation optical CT scans were acquired with the Duke Large field of view Optical CT Scanner (DLOS) and dose was reconstructed with 0.5mm isotropic voxel size. The measured radial dose factor matched the published fit to within 3% for radii between 0.5-3.0cm, and the anisotropy function matched to within 4% except for θ near 0° and 180° and radii >3cm. Further improvements in measurement accuracy may be achieved by optimizing dose, using the high dynamic range scanning capability of DLOS, and irradiating multiple dosimeters. Initial simulations indicate an 8 fold increase in dose is possible while still allowing sufficient light transmission during optical CT. A more comprehensive measurement may be achieved by increasing dosimeter size and flipping the source orientation between irradiations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An Investigation into the Robustness of Optical-CT Dosimetry of a Radiochromic Dosimeter Compatible with the RPC Head-and-Neck Phantom.

The potential of the PRESAGE™/Optical-CT system as a comprehensive 3D dosimetry tool has been demonstrated. The current study focused on detailed characterization of robustness (intra-dosimeter uniformity and temporal stability) and reproducibility (inter-dosimeter reproducibility) of PRESAGE™ inserts compatible with the RPC H&N phantom. In addition, the accuracy and precision of PRESAGE dose m...

متن کامل

A practical three-dimensional dosimetry system for radiation therapy.

There is a pressing need for a practical three-dimensional (3D) dosimetry system, convenient for clinical use, and with the accuracy and resolution to enable comprehensive verification of the complex dose distributions typical of modern radiation therapy. Here we introduce a dosimetry system that can achieve this challenge, consisting of a radiochromic dosimeter (PRESAGE) and a commercial optic...

متن کامل

Investigating the Feasibility of 3D Dosimetry in the RPC IMRT H&N Phantom.

An urgent requirement for 3D dosimetry has been recognized because of high failure rate (~25%) in RPC credentialing, which relies on point and 2D dose measurements. Comprehensive 3D dosimetry is likely to resolve more errors and improve IMRT quality assurance. This work presents an investigation of the feasibility of PRESAGE/optical-CT 3D dosimetry in the Radiologic Physics Center (RPC) IMRT H&...

متن کامل

On the feasibility of polyurethane based 3D dosimeters with optical CT for dosimetric verification of low energy photon brachytherapy seeds.

PURPOSE To investigate the feasibility of and challenges yet to be addressed to measure dose from low energy (effective energy <50 keV) brachytherapy sources (Pd-103, Cs-131, and I-125) using polyurethane based 3D dosimeters with optical CT. METHODS The authors' evaluation used the following sources: models 200 (Pd-103), CS-1 Rev2 (Cs-131), and 6711 (I-125). The authors used the Monte Carlo r...

متن کامل

Optical Characterization of NIPAM and PAGAT Polymer Gels for Radiation Dosimetry

Introduction The purpose of the current study was to determine optical sensitivity of N-isopropyl acrylamide NIPAM and polyacrylamide gelatin and tetrakis hydroxymethyl phosphoniun chloride (PAGAT) polymer gels for different wavelength of visible light spectrum applied in optical computed tomography method. Materials and Methods NIPAM and PAGAT polymer gels with conventional formulations used f...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 444  شماره 

صفحات  -

تاریخ انتشار 2013