Using BD Vacutainer CD4 Stabilization Tubes for Absolute Cluster of Differentiation Type 4 Cell Count Measurement on BD FacsCount and Partec Cyflow Cytometers: A Method Comparison Study from Zimbabwe
نویسندگان
چکیده
BACKGROUND Blood collected in conventional EDTA tubes requires laboratory analysis within 48 hours to provide valid CD4 cell count results. This restricts access to HIV care for patients from rural areas in resource-constraint settings due to sample transportation problems. Stabilization Tubes with extended storage duration have been developed but not yet evaluated comprehensively. OBJECTIVE To investigate stability of absolute CD4 cell count measurement of samples in BD Vacutainer CD4 Stabilization Tubes over the course of 30 days. METHODS This was a laboratory-based method comparison study conducted at a rural district hospital in Beitbridge, Zimbabwe. Whole peripheral blood from 88 HIV positive adults was drawn into BD Vacutainer CD4 Stabilization Tubes and re-tested 1, 2, 3, 5, 7, 14 and 30 days after collection on BD FacsCount and Partec Cyflow cytometers in parallel. Absolute CD4 cell levels were compared to results from paired samples in EDTA tubes analysed on BD FacsCount at the day of sample collection (references methodology). Bland-Altman analysis based on ratios of the median CD4 counts was used, with acceptable variation ranges for Limits of Agreements of +/-20%. RESULTS Differences in ratios of the medians remained below 10% until day 21 on BD FacsCount and until day 5 on Partec Cyflow. Variations of Limits of Agreement were beyond 20% after day 1 on both cytometers. Specimen quality decreased steadily after day 5, with only 68% and 40% of samples yielding results on BD FacsCount and Partec Cyflow at day 21, respectively. CONCLUSIONS We do not recommend the use of BD Vacutainer CD4 Stabilization Tubes for absolute CD4 cell count measurement on BD FacsCount or Partec Cyflow due to large variation of results and decay of specimen quality. Alternative technologies for enhanced CD4 testing in settings with limited laboratory and sample transportation capacity still need to be developed.
منابع مشابه
Opportunities to improve storage and transportation of blood specimens for CD4 testing in a rural district in Zimbabwe using BD vacutainer CD4 stabilization tubes: a stability and diagnostic accuracy study
BACKGROUND CD4+ T-cell testing of blood specimens collected in standard EDTA Vacutainer tubes and transported at ambient temperature, must be completed within 48 hours with the BD FACSCount™ flow cytometer, restricting specimen collection in remote clinics with no on-site testing and limited specimen transport services. We conducted a study in Buhera District, Zimbabwe, to assess the stability ...
متن کاملEvaluation of the Partec flow cytometer against the BD FACSCalibur system for monitoring immune responses of human immunodeficiency virus-infected patients in Zimbabwe.
A single-platform volumetric flow cytometer, the Partec Cyflow SL_3, was evaluated against a BD FACSCalibur/Sysmex XT1800i dual platform for measuring CD4(+) lymphocytes, total lymphocytes, and the percentage of CD4 lymphocytes in whole-blood samples for monitoring the immune systems of human immunodeficiency virus (HIV)/AIDS patients. Statistical analyses for precision, correlation, and agreem...
متن کاملEvaluation of a new single-parameter volumetric flow cytometer (CyFlow(green)) for enumeration of absolute CD4+ T lymphocytes in human immunodeficiency virus type 1-infected Thai patients.
Use of the standard dual-platform flow cytometric method for determination of CD4(+) T-lymphocyte counts, which needs both a flow cytometer (FCM) and hematological analyzer, would inevitably lead to increased variability. The development of new single-platform (SP) FCMs that provide direct CD4(+) T-lymphocyte counts for improved assay precision and accuracy have recently attracted attention. Th...
متن کاملThe Partec CyFlow Counter could provide an option for CD4+ T-cell monitoring in the context of scaling-up antiretroviral treatment at the district level in Malawi.
A study was conducted in rural Malawi to verify (a) whether the Partec CyFlow Counter((R)) for CD4+ T-cell lymphocyte counting in HIV-positive individuals could be introduced into a district hospital laboratory and (b) whether it would produce CD4 counts of acceptable quality. CD4+ cell counting was performed using the Partec CyFlow Counter and the results were compared with a reference method ...
متن کاملLaboratory and Field Evaluation of the Partec CyFlow MiniPOC for Absolute and Relative CD4 T-Cell Enumeration
BACKGROUND A new CD4 point-of-care instrument, the CyFlow miniPOC, which provides absolute and percentage CD4 T-cells, used for screening and monitoring of HIV-infected patients in resource-limited settings, was introduced recently. We assessed the performance of this novel instrument in a reference laboratory and in a field setting in Senegal. METHODOLOGY A total of 321 blood samples were ob...
متن کامل