The effect of vacuum mixing and pre-heating the femoral component on the mechanical properties of the cement mantle.

نویسندگان

  • M Baleani
  • E Bialoblocka-Juszczyk
  • G E Engels
  • M Viceconti
چکیده

We investigated the effect of pre-heating a femoral component on the porosity and strength of bone cement, with or without vacuum mixing used for total hip replacement. Cement mantles were moulded in a manner simulating clinical practice for cemented hip replacement. During polymerisation, the temperature was monitored. Specimens of cement extracted from the mantles underwent bending or fatigue tests, and were examined for porosity. Pre-heating the stem alone significantly increased the mean temperature values measured within the mantle (+14.2 degrees C) (p < 0.001) and reduced the mean curing time (-1.5 min) (p < 0.001). The addition of vacuum mixing modulated the mean rise in the temperature of polymerisation to 11 degrees C and reduced the mean duration of the process by one minute and 50 seconds (p = 0.01 and p < 0.001, respectively). In all cases, the maximum temperature values measured in the mould simulating the femur were < 50 degrees C. The mixing technique and pre-heating the stem slightly increased the static mechanical strength of bone cement. However, the fatigue life of the cement was improved by both vacuum mixing and pre-heating the stem, but was most marked (+ 280 degrees C) when these methods were combined. Pre-heating the stem appears to be an effective way of improving the quality of the cement mantle, which might enhance the long-term performance of bone cement, especially when combined with vacuum mixing.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pre-heating of components in cemented total hip arthroplasty.

Fatigue fractures which originate at stress-concentrating voids located at the implant-cement interface are a potential cause of septic loosening of cemented femoral components. Heating of the component to 44 degrees C is known to reduce the porosity of the cement-prosthesis interface. The temperature of the cement-bone interface was recorded intra-operatively as 32.3 degrees C. A simulated fem...

متن کامل

A Single Femoral Component for All Total Hip Replacements Performed by a Trust? Does This Affect Early Clinical and Radiological Outcomes?

Background: Hospitals may be under pressure to implement cost saving strategies regarding prosthesis choice. Thismay involve the use of components which are not the first preference of individual surgeons, or those they have littleexperience with. We aim to examine the effect of standardizing the type of femoral stem used in a single trust, anddetermine whether this is safe practice, particular...

متن کامل

Vacuum-mixing cement does not decrease overall porosity in cemented femoral stems: an in vitro laboratory investigation.

The role of vacuum mixing on the reduction of porosity and on the clinical performance of cemented total hip replacements remains uncertain. We have used paired femoral constructs prepared with either hand-mixed or vacuum-mixed cement in a cadaver model which simulated intra-operative conditions during cementing of the femoral component. After the cement had cured, the distribution of its poros...

متن کامل

The Influence of Heat-Input on Structural and Mechanical Properties of 2205 Duplex Stainless Steel

The influence of heat-input and pre-heat treatment on the structure, mechanical and corrosion behaviors of 2205 duplex stainless steel joint by means of GMAW process was the goal of this study. In this regards, the welding process was done using different heat input in the range of 0.6 to 1.4 kJ/mm and different pre-heating treatments in the range of 25 to 100 oC. The microstructural properties...

متن کامل

The Influence of Heat-Input on Structural and Mechanical Properties of 2205 Duplex Stainless Steel

The influence of heat-input and pre-heat treatment on the structure, mechanical and corrosion behaviors of 2205 duplex stainless steel joint by means of GMAW process was the goal of this study. In this regards, the welding process was done using different heat input in the range of 0.6 to 1.4 kJ/mm and different pre-heating treatments in the range of 25 to 100 oC. The microstructural properties...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 92 3  شماره 

صفحات  -