Positron emission tomography during transcranial magnetic stimulation does not require mu-metal shielding.

نویسندگان

  • Jae Sung Lee
  • Shalini Narayana
  • Jack Lancaster
  • Paul Jerabek
  • Dong Soo Lee
  • Peter Fox
چکیده

Recording brain activity using positron emission tomography (PET) during the stimulation of different parts of the brain by transcranial magnetic stimulation (TMS) permits the mapping of neural connections in the living human brain. However, controversy remains regarding the need for micro-metal shielding of the PET scanner during magnetic stimulation. The aim of this study was to test the effects of magnetic fields generated by TMS on PET data acquisition. With TMS-on and -off in the PET field of view, transmission scans with a (68)Ge/(68)Ga pin source and emission scans with an uniform phantom filled with water and (18)F were acquired. The frequency and intensity of stimulation were set at 3-5 Hz and 70-80% of the maximum output of the stimulator, respectively. The TMS coil was placed at several locations inside the PET gantry, and the main field direction of the TMS coil was varied between parallel and perpendicular orientation to the scanner's axis. Qualitative and quantitative evaluation of the sinograms of transmission PET scans and reconstructed emission images indicated no measurable differences between TMS-on and -off and post-TMS conditions for any position or orientation. The long distance between the TMS coil and the detector block in the PET scanner, as well as the rapid reduction of the magnetic field with distance (3% of maximum field at 10 cm, in air), could explain the lack of TMS interference. The brief duration (approximately 250 micros) of the TMS pulses relative to the total PET acquisition time would also explain the lack of TMS effects. The lack of TMS effects on the PET scanner, as well as PET imaging without any shielding, has been reported by other laboratories.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Imaging the brain before, during, and after transcranial magnetic stimulation.

This article provides a brief overview of current trends in combining neuroimaging and transcranial magnetic stimulation (TMS). First, I outline the utility of magnetic-resonance imaging (MRI) and frameless stereotaxy for planning, monitoring and documenting the location of the TMS coil relative to the subject's brain. Second, I describe two novel methods, based on the combination of TMS with p...

متن کامل

Functional neuroimaging of the baboon during concurrent image-guided transcranial magnetic stimulation

Transcranial magnetic stimulation (TMS) has well-established applications in basic neuroscience and promising applications in neurological and psychiatric disorders. However the underlying mechanisms of TMS-induced alterations in brain function are not well understood. As a result, treatment design parameters are determined ad hoc and not informed by any coherent theory or model. Once the mecha...

متن کامل

The Neural Correlates of Phonological Short-term Memory: A Repetitive Transcranial Magnetic Stimulation Study

Neuropsychological reports and activation studies by means of positron emission tomography and functional magnetic resonance imaging have suggested that the neural correlates of phonological short-term memory are located in the left hemisphere, with Brodmann's area (BA) 40 being responsible for short-term storage, and BA 44 for articulatory rehearsal. However, a careful review of the literature...

متن کامل

A Comparative Study of the Construction of Positron Emission Tomography/Computed Tomography Facilities in Three South African Hospitals

Introduction: Development of higher energy modalities such as positron emission tomography/computed tomography (PET/CT), has led to more complex shielding problems. This is due to several factors, such as the radiopharmaceutical relatively high-administered activity, high patient throughput, and high energies of 511 kilo-electron volt (keV) positron annihilation photon...

متن کامل

Transcranial magnetic stimulation during positron emission tomography: a new method for studying connectivity of the human cerebral cortex.

We describe a new technique permitting the mapping of neural connections in the living human brain. The method combines two well established tools of brain research: transcranial magnetic stimulation (TMS) and positron emission tomography (PET). We use TMS to stimulate directly a selected cortical area while simultaneously measuring changes in brain activity, indexed by cerebral blood flow (CBF...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • NeuroImage

دوره 19 4  شماره 

صفحات  -

تاریخ انتشار 2003