Hematocrit interference of blood glucose meters for patient self-measurement.
نویسندگان
چکیده
BACKGROUND Abnormal hematocrit levels may interfere with glucose readings of patient self-assessment blood glucose (BG) meters. The aim of this laboratory investigation was to assess the potential influence of hematocrit variations on a variety of BG meters applying different measurement technologies. METHODS Venous heparinized blood was manipulated to contain three different BG concentrations (50-90, 120-180, and 280-350 mg/dl) and five different hematocrit levels (25%, 35%, 45%, 55%, and 65%). After careful oxygenation to normal blood oxygen pressure (65-100 mmHg), each sample was measured (eight times) with the following devices: Accu-Chek® Aviva Nano and Active, Breeze®2 and Contour®, FreeStyle Freedom Lite®, GlucoDr. auto™, Glucofix® mio Plus, GlucoLab™, GlucoMen® LX Plus, Nova Max® Link, Nova Max® Plus, OneTouch® Ultra®2 and Verio®, On Call® Plus and Platinum, Optium Xceed®, Precision Xceed®, and TaiDoc Fora TD-4227. A YSI 2300 STAT Plus™ glucose analyzer served as reference method. Stability to hematocrit influence was assumed, with <10% mean glucose result bias between the highest and lowest hematocrit levels. RESULTS Six of the investigated meters showed a stable performance in this investigation: Accu-Chek Active (7%), Glucofix mio Plus (5%), GlucoMen LX Plus (4%), Nova Max Plus (4%), Nova Max Link (7%), and OneTouch Verio (3%). All other meters failed this hematocrit interference test, with FreeStyle Freedom Lite (11%), and On Call Platinum (12%) being the better devices and On Call Plus (68%), GlucoLab (51%), TaiDoc Fora TD-4227 (39%), and Breeze 2 (38%) showing the worst performance. CONCLUSIONS Hematocrit may affect BG meter performance in daily routine. In case of interference, low hematocrit values (<35%) result in too high readings. Our results encourage use of meters that are not affected by hematocrit interference.
منابع مشابه
Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices.
BACKGROUND It has been demonstrated that dynamic electrochemistry can be used to correct blood glucose measurement results for potentially interfering conditions, such as humidity, hematocrit (HCT) variations, and ascorbic acid. The purpose of this laboratory investigation was to assess the potential influence of hematocrit variations on a variety of blood glucose meters applying different meas...
متن کاملEvaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.
BACKGROUND Most glucose meter comparisons to date have focused on performance specifications likely to impact subcutaneous dosing of insulin. We evaluated four hospital-based glucose meter technologies for accuracy, precision, and analytical interferences likely to be encountered in critically ill patients, with the goal of identifying and discriminating glucose meter performance specifications...
متن کاملAnalysis of an electrochemical blood glucose monitoring system with hematocrit compensation: improved accuracy by design.
The article entitled "Hematocrit Compensation in Electrochemical Blood Glucose Monitoring Systems" by Teodorczyk and colleagues in this issue of Journal of Diabetes Science and Technology demonstrates that the OneTouch(®) Verio™ glucose meter meets current regulatory expectations for glucose meter performance and is relatively free from interference by hematocrit. The lack of influence of hemat...
متن کاملBlood glucose meters employing dynamic electrochemistry are stable against hematocrit interference in a laboratory setting.
BACKGROUND Hematocrit (HCT) is known to be a confounding factor that interferes with many blood glucose (BG) measurement technologies, resulting in wrong readings. Dynamic electrochemistry has been identified as one possible way to correct for these potential deviations. The purpose of this laboratory investigation was to assess the HCT stability of four BG meters known to employ dynamic electr...
متن کاملSuboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
OBJECTIVE The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA(1c) and hematocrit were also measured. ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of diabetes science and technology
دوره 7 1 شماره
صفحات -
تاریخ انتشار 2013