Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter

نویسندگان

  • Turki M Almubrad
  • Uchechukwu L Osuagwu
  • Ibrahim AlAbbadi
  • Kelechi C Ogbuehi
چکیده

AIM To compare the precision of the Topcon SP-3000P noncontact specular microscope (NCSM) and the DGH 500 ultrasound pachymeter (USP). METHODS Triplicate measurements of central corneal thickness (CCT) for 100 eyes were taken with an NCSM and a USP in 2 visits separated by 1 week. Repeatability was assessed by computing the differences between all 3 readings from each subject. Coefficients of repeatability and reproducibility were computed. RESULTS MEAN CCT AS MEASURED BY EACH INSTRUMENT WERE: 518.53 ± 34.96 μm (range 417.33-592.67) and 516.94 ± 33.60 μm (range 431.67-582.67) for sessions 1 and 2 respectively, with the NCSM; 546.69 ± 36.62 μm (range 457.33-617.00) and 549.78 ± 35.26 μm (range 454.00-618.67) for sessions 1 and 2 respectively, with the USP. The ultrasound CCT measurements were consistently higher than those obtained with the NCSM in both sessions 28.17 ± 19.20 μm (mean ± SD, session 1) and 32.81 ± 14.04 (mean ± SD, session 2). The repeatability coefficient for the NCSM was better in both sessions than those for USP (±10 μm vs ± 12 μm in session 1 and ±8 μm vs ±10 μm in session 2). The reproducibility coefficient with the NCSM was half that with the USP (±21 μm vs ±41 μm). CONCLUSION The SP-3000P NCSM is a more precise and reproducible instrument for measurement of CCT than the USP, but both instruments are reliable, useful instruments for measuring CCT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Corneal thickness measurements with the Topcon SP-2000P specular microscope and an ultrasound pachymeter.

OBJECTIVE To compare the reproducibility of measurements obtained with a new pachymetry instrument, the Topcon specular microscope (Topcon SP-2000P; Topcon America Corp, Paramus, NJ), with those obtained by ultrasound pachymetry. METHODS Corneal thickness was measured in 40 eyes of 40 patients 3 times each with the Topcon SP-2000P and an ultrasound pachymeter (DGH 500, DGH Technology Inc, Ext...

متن کامل

Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness

BACKGROUND The purpose of this study was to assess the repeatability and limits of agreement of corneal thickness values measured by a high-frequency ultrasound (Artemis-2), hand-held ultrasound pachymeter (DGH-500) and a specular microscope (SP-3000P). METHODS Central corneal thickness (CCT) was analyzed in this prospective randomized study that included 32 patients (18 men and 14 women) age...

متن کامل

Central corneal thickness measurements with different imaging devices and ultrasound pachymetry.

PURPOSE The aim of this study was to compare 4 methods of central corneal thickness (CCT) measurements in terms of their agreement, repeatability, and measurement time. METHODS CCT was measured in 184 eyes of 92 healthy subjects by the same examiner. The methods used were as follows: noncontact specular microscopy (SM; Topcon SP-3000P; Topcon Corporation, Tokyo, Japan), Pentacam rotating Sche...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 5  شماره 

صفحات  -

تاریخ انتشار 2011