Permanently Implanted Transvenous Pacemakers
نویسندگان
چکیده
Three patients with complete heart block had a high resistance to electrical stimulation of the endocardium through transvenous pacemaker systems. One patient died; one needed reoperation for implantation of epicardial electrodes; and the third required a generator of higher current. The present enthusiasm for permanent transvenous implantation of pacemakers must be tempered somewhat by the knowledge that high endocardial resistance to pacing represents another cause of failure. Simple and rapid methods of determining the current necessary to stimulate the endocardium are available and should be used prior to permanent implantation of the generator. If a substantial margin of safety does not exist, one should consider substituting a pulse generator with greater output or an epicardial electrode system.
منابع مشابه
Permanently implanted transvenous pacemakers. Electrical measurements of function.
Three patients with complete heart block had a high resistance to electrical stimulation of the endocardium through transvenous pacemaker systems. One patient died; one needed reoperation for implantation of epicardial electrodes; and the third required a generator of higher current. The present enthusiasm for permanent transvenous implantation of pacemakers must be tempered somewhat by the kno...
متن کاملPermanent ventricular pacemakers. Comparison of transthoracic and transvenous implantation.
In the series of 86 patients with the Stokes-Adams syndrome or with symptomatic bradycardia managed with permanently implanted ventricular pacemakers reported on, 35 patients had primary implantation of epicardial leads at thoracotomy and 51 had transvenous endocardial electrodes passed via the jugular venous system for permanent ventricular pacing. The transvenous method of permanent pacemaker...
متن کاملPermanently Implanted Transvenous
Three patients with complete heart block had a high resistance to electrical stimulation of the endocardium through transvenous pacemaker systems. One patient died; one needed reoperation for implantation of epicardial electrodes; and the third required a generator of higher current. The present enthusiasm for permanent transvenous implantation of pacemakers must be tempered somewhat by the kno...
متن کاملClinical evaluation of myocardial perforation as a complication of permanent transvenous pacemakers.
A simple and reliable radiographic method to aid in the detection of penetration or perforation of the myocardium by permanent transvenous pacemaker electrodes is described. The "epicardial fat-pad sign" wm M y positive in three Usymptomatic" patients in whom electrodes had perforated the myocardium and in eight of 28 "nonsymptomatic" patients with pacemakers implanted ra long arr three years. ...
متن کاملTemporary inhibition of permanently implanted demand pacemakers.
Temporary inhibition of permanently implanted demand pacemakers has been previously described. Demand pacemakers may be inhibited by waving a magnet over the region of the pacemaker generator or by chest wall stimulation. The former may not inhibit most of the bipolar pacemakers, whereas the latter may be time consuming and may casue patient discomfort. Another method is described which utilize...
متن کامل