Diagnostic test for mucopolysaccharidosis. II. Rapid quantification of glycosaminoglycan in urine samples collected on a paper matrix.
نویسندگان
چکیده
The direct 1,9-dimethylmethylene blue (DMB) method for quantifying sulfated glycosaminoglycan (GAG) in urine (Clin Chem 1989; 35:374-9) has been adapted to a convenient means for sample collection and transport as a test to identify individuals with mucopolysaccharidosis (MPS) storage diseases. Results correlated moderately well (r = 0.85) with those of a commonly used, but more laborious, quantitative method. In studying factors to maximize differentiation of pathological from normal values, we found that GAG excretion (expressed as milligrams GAG per gram creatinine) fits a logarithmic function with respect to age and varies markedly below age five years. This must be considered in developing normative values and forming diagnoses. Of 112 separate urine specimens obtained from 41 MPS patients representing the major MPS diseases, glycosaminoglycan excretion by all exceeded that for age-matched normal individuals. The convenience of this method allowed us to establish the first normative values for three-week-old infants (n = 435) found to have a mean glycosaminoglycan excretion of 179 (SD 86.3) mg of GAG per gram of creatinine. This method improves the diagnostic capability for those MPS diseases that have been particularly difficult to identify (Sanfilippo's syndrome and Morquio's syndrome), and may also provide a test for other disorders with previously unrecognized abnormal excretion of glycosaminoglycan (e.g., mucolipidosis and acromesomelic dysplasia). Most importantly, this MPS diagnostic test is unique in its suitability for mass screening of newborn infants.
منابع مشابه
Mutational analysis of ARSB gene in mucopolysaccharidosis type VI: identification of three novel mutations in Iranian patients
Objective(s): Mucopolysaccharidosis VI (MPS VI) or Maroteaux-Lamy syndrome is a rare metabolic disorder, resulting from the deficient activity of the lysosomal enzyme arylsulfatase B (ARSB). The enzymatic defect of ARSB leads to progressive lysosomal storage disorder and accumulation of glycosaminoglycan (GAG) dermatan sulfate (DS), which causes harmful effects on various organs and tissues an...
متن کاملInteractive Effects of the Carbon Paper, Sodium Bicarbonate and Oral Contraceptive Pills on Morphine Urine Test
Abstract Background and objectives: A major problem for labs. esp. medico legal centers is drug test false positive and negative results. Using carbon paper, sodium bicarbonate or oral contraceptive pills (OCP) are common among addict people to make the results negative. Therefore, we decided to evaluate the effect of carbon paper, sodium bicarbonate or OCP on morphine urine test. Material and ...
متن کاملValidation of Urinary Glycosaminoglycans in Iranian patients with Mucopolysaccharidase type I: The effect of urine sedimentation characteristics
OBJECTIVE The first line-screening test for mucopolysaccharidosis is based on measurement of urinary glycosaminoglycans. The most reliable test for measurement of urine glycosaminoglycans is the 1,9-dimethyleneblue colorimetric assay. Biological markers are affected by ethnical factors, for this reason, the World Health Organization recommends that the diagnostic test characteristics should be ...
متن کاملDiagnostic test for mucopolysaccharidosis. I. Direct method for quantifying excessive urinary glycosaminoglycan excretion.
This direct method for quantifying excessive urinary glycosaminoglycan excretion exploits the specific binding of 1,9-dimethylmethylene blue (DMB). The procedure obviates cumbersome and labor-intensive procedures for separating glycosaminoglycans from other constituents of urine. Pediatric pharmaceutical formulations (except heparin), in concentrations expected in urine, do not interfere with s...
متن کاملThe spot test is not a reliable screening procedure for mucopolysaccharidoses.
To check the reliability of the Ames MPS paper spot test, which is based on the Azure A dye, we sent a series of urine samples to three laboratories where the spot test is part of the metabolic screening for mucopolysaccharidoses. In these laboratories false-negative results ranged between 19% and 35% and false-positive results ranged between 12% and 29% of all samples submitted. In contrast, t...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Clinical chemistry
دوره 35 10 شماره
صفحات -
تاریخ انتشار 1989