Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?
نویسندگان
چکیده
OBJECTIVE An endotracheal tube (ETT) imposes work of breathing on mechanically ventilated patients. Using a bellows-in-a-box model lung, we compared the tube compensation (TC) performances of the Nellcor Puritan-Bennett 840 ventilator and of the Dräger Evita 4 ventilator. MEASUREMENTS AND RESULTS Each ventilator was connected to the model lung. The respiratory rate of the model lung was set at 10 breaths/min with 1 s inspiratory time. Inspiratory flows were 30 or 60 l/min. A full-length 8 mm bore ETT was inserted between the ventilator circuit and the model lung. The TC was set at 0%, 10%, 50%, and 100% for both ventilators. Pressure was monitored at the airway, the trachea, and the pleura, and the data were recorded on a computer for later analysis of the delay time, of the inspiratory trigger pressure, and of the pressure-time product (PTP). The delay time was calculated as the time between the start of inspiration and minimum airway pressure, and the inspiratory trigger pressure was defined as the most negative pressure level. The same measurements were performed under pressure support ventilation of 4 and 8 cmH2O. The PTP increased according to the magnitude of inspiratory flow. Even with 100% TC, neither ventilator could completely compensate for the PTP imposed by the ETT. At 0% TC the PTP tended to be less with the Nellcor Puritan-Bennett 840 ventilator, while at 100% TC the PTP tended to be less with the Dräger Evita 4 ventilator. A small amount of pressure support can be equally effective to reduce the inspiratory effort compared with the TC. CONCLUSION Although both ventilators provided effective TC, even when set to 100% TC they could not entirely compensate for a ventilator and ETT-imposed work of breathing. The effect of TC is less than that of pressure support ventilation. Physicians should be aware of this when using TC in weaning trials.
منابع مشابه
Accuracy of automatic tube compensation in new-generation mechanical ventilators.
OBJECTIVE To compare performance of flow-adapted compensation of endotracheal tube resistance (automatic tube compensation, ATC) between the original ATC system and ATC systems incorporated in commercially available ventilators. DESIGN Bench study. SETTING University research laboratory. SUBJECTS The original ATC system, Dräger Evita 2 prototype, Dräger Evita 4, Puritan-Bennett 840. INT...
متن کاملAutomatic tube compensation: is it worthwhile?
The influence of the endotracheal tube (ETT) during weaning has been questioned. Concerns about the work of breathing (WOB) imposed by the ETT have led to the use of various corrective methods and ventilatory modes to overcome ETT resistance. Despite the theoretical advantages of different approaches, such as pressure support (PSV) and/or automatic tube compensation (ATC), no methods appear to ...
متن کاملEvaluation of pressure support ventilation with seven different ventilators using Active Servo Lung 5000.
In modern emergency and critical care, physicians tend to choose the mode of mechanical ventilation based on spontaneous breathing for the purpose of promoting discharge of pulmonary secretion and preventing atelectasis in patients with acute respiratory insufficiency. However, we often observe "differences in recovery" among patients treated using the same PSV settings beyond "differences in i...
متن کاملThe impact of imposed expiratory resistance in neonatal mechanical ventilation: a laboratory evaluation.
BACKGROUND Small endotracheal tubes (ETTs) and neonatal ventilators can impact gas exchange, work of breathing, and lung-mechanics measurements in infants, by increasing the expiratory resistance (R(E)) to gas flow. METHODS We tested two each of the Babylog 8000plus, Avea, Carestation, and Servo-i ventilators. In the first phase of the study we evaluated (1) the imposed R(E) of an ETT and ven...
متن کاملEffect of leak and breathing pattern on the accuracy of tidal volume estimation by commercial home ventilators: a bench study.
BACKGROUND New home ventilators are able to provide clinicians data of interest through built-in software. Monitoring of tidal volume (VT) is a key point in the assessment of the efficacy of home mechanical ventilation. OBJECTIVE To assess the reliability of the VT provided by 5 ventilators in a bench test. METHODS Five commercial ventilators from 4 different manufacturers were tested in pr...