Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery
نویسندگان
چکیده
BACKGROUND Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060-2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis. METHODS A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created. RESULTS The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p < 0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of -0.57, -2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p < 0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias = 1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods. CONCLUSIONS In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI.
منابع مشابه
Comparison of continuous non-invasive finger arterial pressure monitoring with conventional intermittent automated arm arterial pressure measurement in patients under general anaesthesia.
BACKGROUND For a majority of patients undergoing anaesthesia for general surgery, mean arterial pressure (MAP) is only measured intermittently by arm cuff oscillometry (MAPiNIAP). In contrast, the Nexfin(®) device provides continuous non-invasive measurement of MAP (MAPcNIAP) using a finger cuff. We explored the agreement of MAPcNIAP and MAPiNIAP with the gold standard: continuous invasive MAP ...
متن کاملLevel of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children.
BACKGROUND We compared Nexfin non-invasive arterial pressure measurements using a novel small finger cuff with intra-arterial pressure in the paediatric setting in order to establish the level of agreement between both methods. METHODS The study included 41 children aged 2-16 yr admitted for surgery or paediatric intensive care with an intra-arterial catheter as part of standard monitoring. V...
متن کاملComparison of Intra-abdominal Pressure Measurement and Physical Exam for Diagnosis of Surgery Indication in Patients with Abdominal Compartment Syndrome due to Blunt Trauma
Background & Aims: Increase in abdominal pressure can lead to the so-called intra–abdominal compartment syndrome (ACS) that is often observed during the first 24 hours after sever abdominal trauma and surgery. Measurement of the intra abdominal pressure through the bladder as a non-invasive measurement can provide a quick and accurate assessment of abdominal pressure changes. This study was per...
متن کاملContinuous non-invasive finger arterial pressure monitoring reflects intra-arterial pressure changes in children undergoing cardiac surgery.
BACKGROUND Continuous non-invasive measurement of finger arterial pressure (FAP) is a reliable technology in adults. FAP is measured with an inflatable cuff around the finger and simultaneously converted to a reconstructed brachial artery pressure waveform (reBAP) by the Nexfin™ device. We assessed the adequacy of a prototype device (Nexfin-paediatric), designed for a paediatric population, for...
متن کاملEffect of body position and head of bed angle on intra-abdominal and endotracheal tube cuff pressure in mechanically ventilated patients
Introduction: Changing position is one of the typical nursing care in intensive care units. This study aimed to investigate the effect of body positioning on intra-abdominal pressure and endotracheal tube cuff pressure and their relationship with each other in patients undergoing mechanical ventilation Materials and Methods: This was a before-after clinical trial study performed on 70 patients ...
متن کامل