BACKGROUND Significant paravalvular aortic regurgitation (PAR) after transcatheter aortic valve implantation (TAVI) is associated with negative clinical consequences. We hypothesize that increased eccentricity of the aortic annulus is associated with greater PAR. METHODS Patients with severe aortic stenosis underwent multidetector computed tomography (MDCT) before successful TAVI with the Medtronic CoreValve bioprosthesis. The smallest (D(min)) and largest (D(max)) orthogonal diameters in the basal ring of the aortic annulus were determined. We defined circularity of aortic annulus using the eccentricity index (1 - D(min)/D(max)). The primary endpoint was early occurrence of significant PAR, defined as > grade II PAR by postprocedural aortography. RESULTS Eighty-four patients, mean age 83 ± 4 years with a mean aortic valve area of 0.7 ± 0.2 cm² were included. Twenty patients had postprocedural PAR > grade II. Using a receiver operating characteristic (ROC) analysis, eccentricity index correlated with significant PAR (AUC = 0.834; P=.034). A retrospectively determined eccentricity index cut-off of >0.25 was related to significant PAR with a sensitivity of 80%, specificity of 86%, and negative predictive value of 95% (P<.001). On univariate logistic regression, eccentricity index of >0.25 (P<.001) and device implantation depth (P=.015) correlated with significant PAR, while other parameters such as annular calcification and cover index did not. On multivariate analysis including only parameters with P<.1 on univariate analysis, eccentricity index >0.25 was the sole independent predictor of significant PAR. CONCLUSION Eccentricity index is related to significant PAR after TAVI with Medtronic CoreValve. Further larger studies are required to determine the utility of this novel index in screening suitable patients for this procedure.
منابع مشابه
"Balloon withdrawal technique" to correct prosthesis malposition and treat paravalvular aortic regurgitation during TAVI.
Transcatheter aortic valve implantation (TAVI) is an emerging technology used to treat high-risk patients with severe aortic stenosis. During TAVI with the CoreValve ReValving System, a balloon is used for the reduction of paravalvular regurgitation. However, in this paper, we describe the "balloon withdrawal" technique through which the positioning of a second valve can be avoided in case of i...
متن کاملFDG-PET/CT for driveline infection in a patient with implantable left ventricular assist device.
can Society of Echocardiography Council for Intraoperative Echocardiography and the Society of Cardiovascular Anesthesiologists Task Force for Certification in Perioperative Transesophageal Echocardiography. J Am Soc Echocardiogr 1999;12: 884–900. 21. Wong DT, Bertaso AG, Liew GY, Thomson VS, Cunnington MS, Richardson JD et al. Relationship of aortic annular eccentricity and paravalvular regurg...
متن کاملRelationship of Annular Sizing Using Multidetector Computed Tomographic Imaging and Clinical Outcomes After Self-Expanding CoreValve Transcatheter Aortic Valve Replacement.
BACKGROUND Multidetector computed tomography is useful for determining the appropriate transcatheter heart valve (THV) size in patients with severe aortic stenosis who are suboptimal surgical candidates. The relationship between adherence to the recommended CoreValve sizing algorithm and clinical outcomes is not known. METHODS AND RESULTS We evaluated 1023 patients with severe aortic stenosis...
متن کاملTranscatheter Aortic Valve Replacement: A Review Article
Transcatheter aortic valve replacement (TAVR) is a novel therapeutic intervention for the replacement of severely stenotic aortic valves in high-risk patients for standard surgical procedures. Since the initial PARTNER trial results, use of TAVR has been on the rise each year. New delivery methods and different valves have been developed and modified in order to promote the minimally invasive p...
متن کاملQuantitative analysis of post-TAVI aortic regurgitation with cardiovascular magnetic resonance and the relationship to transthoracic echocardiography
Background Transcatheter Aortic Valve Implantation (TAVI) is increasingly used to treat patients with severe aortic stenosis at high surgical risk. The severity of post-implantation valvular or paravalvular regurgitation has been shown to adversely affect patient outcome. The aim of the study was to assess the prevalence and severity of aortic regurgitation (AR) at 6 months post-TAVI using card...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of invasive cardiology
دوره 25 4 شماره
صفحات -
تاریخ انتشار 2013