Direct detection of 6 MV x-rays from a medical linear accelerator using a semiconducting polymer diode.
نویسندگان
چکیده
Recently, a new family of low-cost x-radiation detectors have been developed, based on semiconducting polymer diodes, which are easy to process, mechanically flexible, relatively inexpensive, and able to cover large areas. To test their potential for radiotherapy applications such as beam monitors or dosimeters, as an alternative to the use of solid-state inorganic detectors, we present the direct detection of 6 MV x-rays from a medical linear accelerator using a thick film, semiconducting polymer detector. The diode was subjected to 4 ms pulses of 6 MV x-rays at a rate of 60 Hz, and produces a linear increase in photocurrent with increasing dose rate (from 16.7 to 66.7 mGy s(-1)). The sensitivity of the diode was found to range from 13 to 20 nC mGy(-1) cm(-3), for operating voltages from -50 to -150 V, respectively. The diode response was found to be stable after exposure to doses up to 15 Gy. Testing beyond this dose range was not carried out. Theoretical calculations show that the addition of heavy metallic nanoparticles to polymer films, even at low volume fractions, increases the x-ray sensitivity of the polymer film/nanoparticle composite so that it exceeds that for silicon over a wide range of x-ray energies. The possibility of detecting x-rays with energies relevant to medical oncology applications opens up the potential for these polymer detectors to be used in detection and imaging applications using medical x-ray beams.
منابع مشابه
The microscopic dose enhancement factor measurment of bismuth nanoparticles in external radiation therapy using MRI polymer gel dosimetery
Introduction: Bismuth-based nanoparticles having high atomic number and electron density, besides of being utilized as CT contrast agent, stand as a strongly proposed method for dose enhancement in radiotherapy. Embedded nanoparticle in polymer gel dosimeter poses the first step to assess the dose enhancement capability in nanoparticles. To this end, MAGICA gel dosimeter, whic...
متن کاملMeasurement of stereotactic beam data of 9 MV X-rays using diode detector
Background: Radiosurgery is a focal brain irradiation technique that delivers, usually in a single session, high dose of ionizing radiation. The presence of lateral electronic disequilibrium and steep dose gradients in small fields demands special attention to the selection of a suitable detector with respect to its size, composition and response. Materials and Methods : Small circular fields w...
متن کاملCharacterization of thick film poly„triarylamine... semiconductor diodes for direct x-ray detection
Thick film 5 m thick semiconducting polymer diodes incorporating poly triarylamine PTAA have been produced and applied as direct x-ray detectors. Experiments determined that a rectifying diode behavior persists when increasing the thickness of the active layer above typical thin film thicknesses 1 m , and the electrical conduction mechanism of the diodes has been identified. Direct current and ...
متن کاملThe dosimetry assessment of Varian Linear Accelerators of 6, 15 and 20 MV by Monte Carlo Method
Introduction: Monte Carlo method is often applied in radiation therapy as utilized in all the branches of science. An important requirement for successful radiotherapy is carefully examine the dose distribution specifications and decrease the difference between these features with experience to an acceptable level. In this study, the characteristics of 6, 15 and 20 MeV incident x-rays are prov...
متن کاملFlexible radiation dosimeters incorporating semiconducting polymer thick films
Flexible radiation dosimeters have been produced incorporating thick films (>1 μm) of the semiconducting polymer poly([9,9-dioctylfluorenyl-2,7-diyl]-co-bithiophene). Diode structures produced on aluminium-metallised poly(imide) substrates, and with gold top contacts, have been examined with respect to their electrical properties. The results suggest that a Schottky conduction mechanism occurs ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Physics in medicine and biology
دوره 58 13 شماره
صفحات -
تاریخ انتشار 2013