Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan.
نویسندگان
چکیده
PURPOSE The purpose of this study was to determine the validity and repeatability of anterior chamber depth (ACD) measurements obtained with a novel rotating Scheimpflug camera (Pentacam; Oculus Optikgeräte GmbH, Wetzlar, Germany) to scanning slit topography (Orbscan; Bausch & Lomb, Rochester, NY). METHODS ACD in 60 healthy eyes was measured twice by two independent observers with each modality in random order. A total of eight measurements was performed on each eye. The mean differences between measurements, between observers, and between modalities were calculated, and 95% limits of agreement (LoA) were given as mean +/- 1.96 * standard deviation (SD) of the mean. RESULTS The mean ACD values as determined with the different modalities (+/- SD) were 3.18 +/- 0.38 mm for Pentacam imaging and 3.23 +/- 0.40 for Orbscan. The mean difference between the two modalities was 0.047 mm (LoA: 0.176 to -0.081 mm). Within each modality, the limits of agreement of the differences between individual measurements and between different observers were within 0.073 mm or 4% of total ACD. The total variance in the observed data was 0.145. The variance resulting from interindividual differences in ACD was 85%, as a result of the two different modalities 15%, as a result of different observers 0.007%, and as a result of different measurements 0.007%. CONCLUSIONS In the assessment of normal eyes, the differences of ACD values measured with Orbscan and Pentacam were within clinically acceptable levels, and inter- and intraobserver variability was considerably below clinically significant levels. Thus, these two modalities can be regarded as interchangeable.
منابع مشابه
Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes.
PURPOSE This study determined to assess the degree of agreement between anterior chamber depth (ACD) measurements obtained using three different devices and to analyse the relationship between ACD and spherical equivalent (SE) refraction. METHODS In this cross-sectional study, 42 eyes of 42 patients with a mean SE of - 4.69 +/- 4.61 D (range 0.00 D to - 14.88 D) were analysed. Measurements of...
متن کاملComparison of three methods of measuring corneal thickness and anterior chamber depth.
PURPOSE To compare three different methods of measuring corneal thickness (CT) and anterior chamber depth (ACD). DESIGN Prospective clinical trial (Medical University of Vienna, Austria). METHODS Central CT (CCT), CT at four peripheral points, and central ACD were measured in 88 eyes of 44 healthy subjects with the Pentacam (rotating Scheimpflug camera; Oculus, Wetzlar, Germany), Orbscan I ...
متن کاملEfficacy of Orbscan II and Pentacam topographers by a repeatability analysis when assessing elevation maps in candidates to refractive surgery.
INTRODUCTION Anterior and posterior corneal elevations are measurements used to detect keratoconus suspects. PURPOSE To determine the efficacy of Orbscan II and Pentacam when assessing their elevation maps. MATERIALS AND METHODS The efficacy of the Orbscan II and Pentacam measuring the anterior and posterior corneal elevations were evaluated in a sample of 68 eyes. The concordance between t...
متن کاملComparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy
PURPOSE The aim of this study was to determine the agreement between Pentacam HR (Scheimpflug imaging, Oculus) and Orbscan II (scanning slit topography, Bausch and Lomb) in measuring corneal parameters after photorefractive keratectomy (PRK) for hyperopia. METHODS In this prospective cross-sectional study, 38 hyperopic eyes undergoing PRK were examined before refractive surgery and 8 to 10 mo...
متن کاملComparison of anterior chamber depth measurement between Orbscan IIz and ultrasound biomicroscopy.
PURPOSE To compare the measurement of anterior chamber depth exclusive of corneal thickness using Orbscan IIz and ultrasound biomicroscopy (UBM) and evaluate the repeatability of each method. METHODS Three consecutive measurements of anterior chamber depth were prospectively performed using Orbscan IIz and Paradigm 50-MHz UBM in 40 eyes in 20 individuals. Mean values were compared using the p...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Optometry and vision science : official publication of the American Academy of Optometry
دوره 82 9 شماره
صفحات -
تاریخ انتشار 2005