Troponin-specific autoantibody interference in different cardiac troponin I assay configurations.
نویسندگان
چکیده
BACKGROUND Autoantibodies to cardiac troponins (cTnAAb) can interfere with the measurement of cardiac troponin I (cTnI) by immunoassays. The aim of this study was to explore the degree of cTnAAb interference in different cTnI assay configurations. METHODS Ternary troponin complex was added into samples (serum or plasma, n = 132, 68% cTnAAb positive) from individuals without known cardiac conditions. The recovery of cTnI was then measured with 6 investigational cTnI assays (2, 3, or 4 antibodies per assay). Three of these assays were then selected for further comparison by use of samples (plasma, n = 210, 33% cTnAAb positive) from non-ST-elevation acute coronary syndrome patients in the FRISC-II (FRagmin/Fast Revascularisation during InStability in Coronary artery disease) cohort. Finally, these results were compared to those obtained with 3 commercial cTnI assays. RESULTS Analytical recoveries varied widely among the 6 investigational assays. Notably the low recoveries (median 9%) of the midfragment-targeting reference assay were normalized (median 103%) with the use of the 4-antibody assay construct (3 capture, 1 tracer antibody) with only 1 antibody against a midfragment epitope. Reduced analytical recoveries correlated closely with measured autoantibody amounts. cTnI concentrations from cTnAAb-positive patient samples determined with 3 investigational assays confirmed the reduced concentrations expected from the low analytical recoveries. The results from the commercial cTnI assays with antibody selections representative for contemporary assay constructs revealed a similar underestimation (up to 20-fold) of cTnI in cTnAAb-positive samples. CONCLUSIONS A novel cTnI assay deviating from the conventional IFCC-recommended midfragment approach substantially improves cTnI detection in samples containing cTnAAbs.
منابع مشابه
Cardiac troponin T quantitative assay failure as a result of antibody interference
BACKGROUND Immunoassays are prone to interference by various substances which may cause inaccurate results. This type of interference is difficult to detect analytically. OBJECTIVE A case of CARDIAC Troponin T Quantitative reader (Roche Diagnostics) assay failure was detected and investigated in order to ascertain the likely cause. METHOD Patient whole blood was mixed with cardiac troponin ...
متن کاملAutoantibodies to cardiac troponin associate with higher initial concentrations and longer release of troponin I in acute coronary syndrome patients.
BACKGROUND Cardiac troponin (cTn) is an established marker of myocardial infarction. Pronounced heterogeneity and the minute amounts released into the circulation constitute significant challenges for cTn detection. Recently, autoantibody formation to cTn was shown to be common and to interfere with immunoassay performance. In this study, we investigated cTn autoantibodies and cardiac troponin ...
متن کاملAnalysis of the relation between coping ways with stress and Cardiac Biomarker Troponin I in coronary heart disease patients
Introduction: Coping ways with stress in coronary heart disease patients can lead to significant changes in levels of Cardiac Biomarker Troponin I, but researchers had not explored it empirically. So, the main objective of this study was to identify the relationship between coping ways with stress and Cardiac Biomarker Troponin I and also, finding coping ways that predict changes in the concent...
متن کاملسنجش مقادیر تروپونین ? قلبی در بیماران مبتلا به آمبولی حاد ریه و ارتباط آن با اختلال عملکرد بطن راست
Background and purpose: Right ventricular dysfunction is common in major pulmonary embolisms. The purpose of this study was to evaluate the prevalence and diagnostic utility of cardiac troponin I and also to identify patients with RV dysfunction in pulmonary embolism. Materials and methods: This study was conducted on 42 patients with pulmonary embolism in Ekbatan Hospital, in Hamedan city. ...
متن کاملSelection of antibodies and epitopes for cardiac troponin immunoassays: should we revise our evidence-based beliefs?
Cardiac troponins are regarded as the most specific and sensitive of the currently available diagnostic techniques for detection of myocardial damage, and the redefined criteria used to classify acute coronary syndrome patients presenting with ischemic symptoms as myocardial infarction patients are heavily predicated on an increased concentration of these markers in blood (1 ). Troponins also a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 58 6 شماره
صفحات -
تاریخ انتشار 2012