First-third ejection fraction: is the first-pass radionuclide method accurate?

نویسندگان

  • R N Pierson
  • A Yamashina
  • M I Friedman
  • N Roistacher
چکیده

Previous reports have suggested that left ventricular first-third ejection fraction (EF) can be obtained from the left ventricular time-activity curve derived from first-pass radionuclide angiography based on Anger camera data. The validity of this technique was assessed by: a study of beat-to-beat variations in data from 15 patients in which electrocardiographic data were simultaneously recorded, and a computer simulation incorporating the application of Poisson statistics to appropriate count rate data. The results of patients studies showed no consistent trend in any first-third parameter obtained from consecutive beats in individual subjects, and unacceptably high statistical uncertainty in the calculation of the first-third ejection fraction. The weighted standard deviation of the first-third ejection fraction in each of 15 patients studied averaged 7.5 EF units, while first-third ejection fraction averaged 22.9 EF units. The relative error averaged 32%. The computer simulation indicated a high relative error of 47% associated with the first-third ejection fraction at typical end-diastolic count rates of 200 per frame from 1,000 computer Poisson randomizations of an appropriate analog volume curve. The results render the first-pass radiocardiographic method invalid for determining first-third ejection fraction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Non-invasive measurement of left ventricular function in coronary artery disease. Comparison of first pass radionuclide ventriculography, M-mode echocardiography, and systolic time intervals.

Fifty consecutive patients having had cardiac catheterisation for coronary artery disease also underwent testing by three non-invasive methods commonly employed for assessment of left ventricular function. These included the first pass radionuclide ejection fraction, fractional shortening of the M-mode echocardiographic left ventricular internal dimension, and pre-ejection period/left ventricul...

متن کامل

Left ventricular diastolic performance at rest and during exercise in patients with coronary artery disease. Assessment with first-pass radionuclide angiography.

We used first-pass radionuclide angiocardiography to assess filling fraction during the first third of diastole, peak filling rate and peak filling rate during the first third of diastole as indexes of left ventricular diastolic performance at rest and after upright bicycle exercise in 32 normal patients and 68 patients with coronary artery disease. The mean filling fraction was unchanged from ...

متن کامل

Use of first-pass radionuclide ventriculography in assessment of wall motion abnormalities induced by incremental atrial pacing in patients with coronary artery disease.

In order to show that stress-induced changes in ventricular function can be determined by radionuclide techniques, 13 patients with coronary artery disease were paced to angina. Changes in ejection fraction and regional wall motion were measured from contrast angiograms and first-pass nuclear angiograms performed using a multicrystal gamma camera. A count-volume method applied to the nuclear te...

متن کامل

The reliability of measuring left ventricular ejection fraction by radionuclide cardiography: evaluation by the method of variance components.

A statistical model based on the method of variance components was applied to obtain confidence statements for single and repeat determinations of left ventricular ejection fraction by radionuclide techniques. With this approach variance caused by individual factors in the measurement procedure is estimated to allow calculation of confidence intervals based on single measurements and the detect...

متن کامل

Left Ventricular Diastolic Performance

We used first-pass radionuclide angiocardiography to assess filling fraction during the first third of diastole, peak filling rate and peak filling rate during the first third of diastole as indexes of left ventricular diastolic performance at rest and after upright bicycle exercise in 32 normal patients and 68 patients with coronary artery disease. The mean filling fraction was unchanged from ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of nuclear medicine : official publication, Society of Nuclear Medicine

دوره 26 9  شماره 

صفحات  -

تاریخ انتشار 1985