limb lengthening using the precicetm nail system: complications and results

نویسندگان

ulrich wiebking department of trauma, medical school hannover, hannover, germany; department of trauma, medical school hannover, hannover, germany. tel: +49-5115322050, fax: +49-5115325877

emmanouil liodakis department of trauma, medical school hannover, hannover, germany

mohamed kenawey department of orthopedic surgery, sohag university, sohag, egypt

christian krettek department of trauma, medical school hannover, hannover, germany

چکیده

conclusions we report of loss of achieved length after lengthening with a telescopic nail. weight bearing before complete consolidation of the regenerate might be a risk factor for that. thorough examination of the limb length and careful evaluation of the radiographs are required in the follow-up period. the precice nail system requires the same vigilance like the other intramedullary systems too. objectives the aim of the current study was to describe and analyze the complications associated with lengthening with the precicetm nail. are the problems preventable when using the precice, related to the distraction rate control, the lengthening goals and technique and handling? methods we retrospectively reviewed the charts of 9 patients operated between 2012 and 2013 with a precicetm nail for a leg length discrepancy (lld). the mean age of the patients was 32 years (range, 17 - 48 years). there were 5 femoral and 4 tibial procedures. the causes of lld were posttraumatic (n = 5) and congenital (n = 4). the mean lld was 36.4 ± 11.4 mm. the minimum follow-ups were 2 months (average, 5 months; range, 2 - 9 months). results the mean distraction rate was 0.5 ± 0.1 mm/day. we observed in 7 patients differences in achieving the lengthening goals (average, 1.6 mm; range, -20.0 - 5.0 mm). average lengthening was 34.7 ± 10.7 mm. all patients reached normal alignment and normal joint orientation. an unintentional loss of the achieved length during the consolidation phase was noticed in patients with delayed bone healing in two cases. in the first case (loss of 20mm distraction) the nail could be redistracted and the goal length was achieved. in the second case (loss of 10mm distraction) the nail broke shortly after the diagnosis and the nail was exchanged. background three types of telescopic nails are mainly used for intramedullary limb lengthening nowadays. despite some important advantages of this new technology (e.g. controlled distraction rate, not restricted availability, possibility to perform accordion maneuvers), few articles exist on clinical results and complications after lengthening with the precicetm nail (ellipse, usa).

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Limb Lengthening Using the PRECICETM Nail System: Complications and Results

BACKGROUND Three types of telescopic nails are mainly used for intramedullary limb lengthening nowadays. Despite some important advantages of this new technology (e.g. controlled distraction rate, not restricted availability, possibility to perform accordion maneuvers), few articles exist on clinical results and complications after lengthening with the PRECICETM nail (Ellipse, USA). OBJECTIVE...

متن کامل

The results of limb lengthening by callus distraction using an extending intramedullary nail (Fitbone) in non-traumatic disorders.

Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (...

متن کامل

PRECICE intramedullary limb lengthening system.

The PRECICE(®) Intramedullary Limb Lengthening System (Ellipse Technologies Inc., CA, USA) is a remotely controlled, magnetically driven, implantable limb lengthening intramedullary nail system. It has both CE mark and US FDA clearance for its first- (2011) and second-generation (2013) implants. It is indicated for the treatment of limb length discrepancy and short stature. It has been used wor...

متن کامل

Clinical Results and Complications of Lower Limb Lengthening for Fibular Hemimelia

Fibular hemimelia is a rare but the most common congenital long bone deficiency, encompassing a broad range of anomalies from isolated fibular hypoplasia up to substantial femoral and tibial shortening with ankle deformity and foot deficiency. Most cases of fibular hemimelia manifest clinically significant leg length discrepancy (LLD) with time that requires adequate correction by bone lengthen...

متن کامل

Limb Lengthening by Implantable Limb Lengthening Devices

Implantable limb lengthening using noninvasively adjusted telescopic nails dates back to 1983. The newest technology is the Precice (Ellipse Technologies). A retrospective study of the first 65 Precice nails was carried out for the treatment of limb length discrepancy (unilateral) and short stature (bilateral). Successful lengthening was achieved in all patients. There were numerous distraction...

متن کامل

Lengthening and deformity correction about the knee using a magnetic internal lengthening nail

INTRODUCTION The introduction of the internal lengthening nail (ILN) has changed the treatment of complex malalignment and shortening about the knee. Acute correction of the deformity and gradual lengthening through this osteotomy site has greatly simplified postoperative recovery. This manuscript is a review of the techniques that are currently being used in surgery. METHODS The article is b...