Light transmission of phototherapy eyeshields.

Twelve types of phototherapy eyeshields showed peak light transmission of less than 0.1%, and none transmitted greater than 0.04% light in the 460 nm spectral region. Commercial eyeshields offered no advantage over locally made ones. The choice of shield may be less important than how it is secured over the infant's eyes.

متن کامل

Light transmission measurements and phototherapy eyepatches.

The transmission characteristics of phototherapy eyeshields have been measured under conditions that mimic the clinical situation. Peak transmission (less than 10%) was detected at 700 nm for the poorest patch tested and less than 2% for either of the other patches examined. The optics of measurement systems are considered with reference both to this and previous studies of light transmission t...

متن کامل

Uneven distribution of light in standard phototherapy.

Maldistribution of light is a potential cause of phototherapy failure in the neonate. Differences in light intensity over the surface can be measured with a radiometer as this detects the radiant energy impinging on the incubator mattress. An infant nursed on the front third of the mattress receives less than 40% of the maximal light incident on the centre of the mattress, and consequently degr...

متن کامل

Phototherapy with turquoise versus blue light.

Preterm jaundiced infants were treated by phototherapy with a new turquoise fluorescent lamp. This was more effective in reducing plasma total bilirubin in relation to light irradiance than the ubiquitously used blue fluorescent lamp.

متن کامل

Light Transmission through Three Types of Translucent Zirconia

  Objective: Zirconia cores have limited light transmittance and data are scarce on light transmission through zirconia cores with and without the veneering ceramic.   Methods: In this in vitro study, Disc-shaped specimens (11.5 mm in diameter and 0.4 (0.05) mm in thickness) were fabricated of three types of zirconia namely Mamut, Heany and ZirkonZahn (n=5). A disc-shaped specimen (11.5 mm in d...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ژورنال

عنوان ژورنال: Archives of Disease in Childhood

سال: 1987

ISSN: 0003-9888,1468-2044

DOI: 10.1136/adc.62.9.970