Fracture of Revision Modular Femoral Stem

The number of revision total hip arthroplasties is predicted to more than double in USA over the next 25 years from an estimated 40800 in 2005 to approximately 96700 in 2030 [1]. Modular femoral stems offer the advantage of adjustment and restoration of joint kinematics including leg length, femoral version and offset, regardless of the exact position of the distal part of the stem [2]. Disadva...

متن کامل

Fracture of cementless femoral stems at the mid-stem junction in modular revision hip arthroplasty systems.

BACKGROUND Mechanical failure of femoral stems at the modular junction of revision hip arthroplasty systems has been reported only infrequently. In the current study, the cause of six stem fractures, which occurred in vivo, was analyzed with use of clinical data and failure analysis. METHODS Six patients with a fracture at the mid-stem junction of a modular revision hip implant were identifie...

متن کامل

Revision of failed total hip arthroplasty with a proximal femoral modular cemented stem.

We have designed a modular cemented femoral component for revision of failed total hip arthroplasty in which deficiency of the proximal femur is such as to require a variable extrafemoral portion of the stem. We present the results of the first 74 operations in 72 patients; 56 of the patients had grade-3 or grade-4 femoral deficiency as defined by Gustilo and Pasternak and 24 had fractures of t...

متن کامل

Early fracture of the modular neck of a MODULUS femoral stem

We present the case of a 46-year-old woman who underwent revision surgery approximately 4 years after total hip arthroplasty because of a fracture of the modular neck of a MODULUS femoral stem. The fractured surfaces of the retrieved implant were inspected using optical and scanning electron microscopy. Three-dimensional finite element analysis was also performed to identify the stresses that m...