Cross-Calibration of Bone Mineral Densities and Body Composition between GE Lunar Prodigy and Osteosys Primus
نویسندگان
چکیده
Background: The aim of this study was to investigate the correlation between bone mineral density (BMD) and body composition measured by Osteosys Primus® GE Lunar Prodigy® calculate conversion rate 2 devices.Methods: 40 subjects were men women in aged 20 29 years old. All participants scanned twice on both Primus (OsteoSys) Prodigy (GE Healthcare) DXA systems using manufacturers’ standard scanning positioning protocols.Results: Compared device, mean fat mass overestimated be 12.1% (1,776.9 g) whole body, 5.1% (163.9 gynoid, 6.7% (87.2 android. with BMDs underestimated 2.3% (0.023 g/cm2) 3.1% (0.035 L1-4. lean derived (1,045.3 total 3.8% (179.4 arms, 7.7% (1,104.8 legs, respectively. There a strong BMD groups.Conclusions: Linear correction equations developed ensure comparability muscle Prodigy. Importantly, use from previous studies would have increased discrepancy
منابع مشابه
Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements
In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21-72 years) underwent anteroposterior lumbar spine L2-L4 (n = 72), proximal femur (n = 72), and total body (n = 62) measurements with th...
متن کاملA comparison of bone mineral densities and body composition between Southeast Asia college students and Chinese college students
The aim of this study was to compare bone mineral densities (BMDs) and body composition between Southeast Asia college students and Chinese college students, in order to provide a certain reference enhancing college students' physical fitness.A total of 1694 Chinese college students (294 men and 1400 women, aged 18-22 years) and 250 Southeast Asia college students (148 men and 102 women, aged 1...
متن کاملBone mineral densities in normal Bangladeshi women
fracture. Several chronic medical conditions, medications, and life style factors affect bone mineral accrual. This study aims to evaluate the bone mineral densities and hip axis lengths in the local population. Materials and Methods: 336 normal Bangladeshi women age ranges 20 to 70 years had evaluation of their bone mineral densities (BMDs) by means of dual X-ray absorptiometry (DXA). Results:...
متن کاملcohesion and cohesive devices in a contrastive analysis between ge and esp texts
the present study was an attempt to conduct a contrastive analysis between general english (ge) and english for specific purposes (esp) texts in terms of cohesion and cohesive devices. to this end, thirty texts from different esp and ge textbooks were randomly selected. then they were analyzed manually to find the frequency of cohesive devices. cohesive devices include reference, substitution, ...
15 صفحه اولRelationship between body composition and bone mineral content in young and elderly women.
PRIMARY OBJECTIVE To study the relationship between bone mineral content (BMC), lean tissue mass (LTM) and fat mass (FM) in a large sample of young and elderly women. RESEARCH DESIGN Cross-sectional. METHODS AND PROCEDURES BMC, LTM and FM were measured by dual-energy X-ray absorptiometry in 2009 free-dwelling Caucasian women aged 63 +/- 7 years (mean +/- SD; range: 37-88 years). The majorit...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ژورنال
عنوان ژورنال: Journal of Bone Metabolism
سال: 2021
ISSN: ['2287-7029', '2287-6375']
DOI: https://doi.org/10.11005/jbm.2021.28.3.215