Comparison of the cycloplegic refractive measurements with handheld, table-mounted refractometers and retinoscopy in children
نویسندگان
چکیده
Background: Handheld autorefractometers are now widely used for screening refractive errors in children. The purpose of the study was to compare measurements from table-mounted, handheld and retinoscopy Material methods: Measurements children with poor visual acuity and/or strabismus were obtained 2WIN table-mounted Nidek ARK-1 refractometers after instillation 0.5% atropine. Data on sphere, spherical equivalent (SE), cylindrical vectors at 0 degrees (J0) 45 (J45) analysed. Results: collected 57 (mean age, 4.3 years ± 2.0 years). refractometer measure statistically significantly lower SE mean values than or (1.67 1.48 D, 2.96 1.95 2.92 1.93 respectively). 95% LOA narrowest SE, J0, J45 vector retinoscopy. difference between more pronounced higher J45. Conclusion: autorefractor provided a reading similar that streak autorefractor. differences other two methods values, so careful interpretation autorefraction results would be advised, especially who most significant risk amblyopia.
منابع مشابه
Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children
PURPOSE To evaluate the performance of the hand-held and table-top autorefractokeratometer in measuring refractive errors by comparing them with cycloplegic retinoscopy. METHODS Included in the study were 112 eyes of 112 pediatric patients whose mean age was 6.78 ± 2.61 years (range, 2 to 12 years). The refractive errors of all the eyes were measured with and without cycloplegia using a hand ...
متن کاملA comparison of cycloplegic refraction to the near retinoscopy technique for refractive error determination.
The near retinoscopy technique of refractive error determination was compared to the standard method of cycloplegic refraction using 10 "infants" (3-12 months of age) and 10 "children" (32-109 months of age). There was a significant difference between the techniques for both sphere and cylinder power. Although there was no interaction of refractive technique and age group, the difference betwee...
متن کاملNon-cycloplegic spherical equivalent refraction in adults: comparison of the double-pass system, retinoscopy, subjective refraction and a table-mounted autorefractor.
AIM To evaluate the accuracy of spherical equivalent (SE) estimates of a double-pass system and to compare it with retinoscopy, subjective refraction and a table-mounted autorefractor. METHODS Non-cycloplegic refraction was performed on 125 eyes of 65 healthy adults (age 23.5±3.0 years) from October 2010 to January 2011 using retinoscopy, subjective refraction, autorefraction (Auto kerato-ref...