zeynab sanjabi m.sc. of radiation medicine dept., shahid beheshti university, tehran, iran. alireza kamali asl assistant professor, nuclear engineering dept., shahid beheshti university, tehran, iran mohammad hassan zahmatkesh associate professor, shahid beheshti university, evin, tehran, iran saeid bagheri m.sc. in medical physics, shahid beheshti university, tehran, iran

introduction: different dosimetric modalities (e.g., ionization chambers, semiconductors, tlds, radiographic films) often have some problems determining isodose curves and percentage depth doses in high dose gradient regions. research has been ongoing to develop portable high resolution dosimetry devices with accurate absolute dosimetry capabilities and easy data analysis. the ‘presage’ dosimet...

Journal: :Journal of physics. Conference series 2009
H S Sakhalkar J Adamovics G Ibbott M Oldham

The potential of the PRESAGE™/Optical-CT system as a comprehensive 3D dosimetry tool has been demonstrated. The current study focused on detailed characterization of robustness (intra-dosimeter uniformity and temporal stability) and reproducibility (inter-dosimeter reproducibility) of PRESAGE™ inserts compatible with the RPC H&N phantom. In addition, the accuracy and precision of PRESAGE dose m...

Alireza Kamali Asl, Mohammad Hassan Zahmatkesh Saeid Bagheri Zeynab Sanjabi

Introduction: Different dosimetric modalities (e.g., ionization chambers, semiconductors, TLDs, radiographic films) often have some problems determining isodose curves and percentage depth doses in high dose gradient regions. Research has been ongoing to develop portable high resolution dosimetry devices with accurate absolute dosimetry capabilities and easy data analysis. The ‘PRESAGE’ dosimet...

2013
T Gorjiara Z Kuncic J Adamovics C Baldock

PRESAGE is a radiochromic solid dosimeter which shows promising potential for 3D proton beam dosimetry. Since an idea dosimeter should be water-equivalent, total depth dose distributions in two PRESAGE formulations irradiated by a 62 MeV proton beam were compared with that in water using GEANT4 Monte Carlo simulations. The dose delivered by secondary particles was also calculated. Our results s...

2013
C - S Wuu Y Xu X Qian J Adamovics E Cascio

Monte Carlo water-equivalence study of two PRESAGE® formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE® is a radiochromic solid dosimeter which shows promising potential for 3D proton beam dosimetry. Since an idea dosimeter should be water-equivalent, total depth dose distributions in two PRESAGE® formulations irra...

2013
T Juang J Newton S Das J Adamovics M Oldham

Deformable 3D dosimeters have potential applications in validating deformable dose mapping algorithms. This study evaluates a novel deformable PRESAGE® dosimeter and its application toward validating the deformable algorithm employed by VelocityAI. The deformable PRESAGE® dosimeter exhibited a linear dose response with a sensitivity of 0.0032 ΔOD/(Gy/cm). Comparison of an experimental dosimeter...

Journal: :Medical physics 2012
M Carroll G Ibbott R Grant J Adamovics M Gillin

PURPOSE To evaluate the effects of overlapping dose volumes for varying field arrangements in two formulations of PRESAGE®: one intended for, and irradiated with, proton beams and the other photon beams. METHODS For each treatment modality (photon, proton), three overlapping field setups were performed. These included a stationary dosimeter irradiated over six fractions, a dosimeter shifted l...

Journal: :Medical physics 2013
M Carroll G Ibbott M Gillin J Adamovics

PURPOSE This investigation studies the feasibility of PRESAGE, a 3D polyurethane dosimeter, for relative dosimetry measurements of clinically relevant proton treatments using an anthropomorphic head phantom developed by the Radiological Physics Center (RPC). Performance of a low-LET dependent PRESAGE was evaluated by comparison to the traditionally used radiochromic film, EBT2, and thermolumine...

Journal: :Medical physics 2006
Pengyi Guo John Adamovics Mark Oldham

There is a pressing need for a practical three-dimensional (3D) dosimetry system, convenient for clinical use, and with the accuracy and resolution to enable comprehensive verification of the complex dose distributions typical of modern radiation therapy. Here we introduce a dosimetry system that can achieve this challenge, consisting of a radiochromic dosimeter (PRESAGE) and a commercial optic...

2010
Cheng-Shie Wuu

Both polymer gels and PRESAGE radiochromic solid dosimeter, in conjunction with optical CT scanning system, have been employed to measure 3-D dose distribution. The 3-D dose maps obtained from these systems can provide a useful tool for dose verification on complex treatments such as IMRT, radiosurgery, and RapidArc. These complex treatments present high dose gradient regions in the boundaries ...