نتایج جستجو برای: imrt qa
تعداد نتایج: 9935 فیلتر نتایج به سال:
PURPOSE The purposes of this study were two-fold: first, to develop a four-axis moving phantom for patient-specific quality assurance (QA) in surrogate signal-based dynamic tumor-tracking intensity-modulated radiotherapy (DTT-IMRT), and second, to evaluate the accuracy of the moving phantom and perform patient-specific dosimetric QA of the surrogate signal-based DTT-IMRT. METHODS The four-axi...
This study compares the EPID dosimetry algorithms of two commercial systems for pretreatment QA, and analyzes dosimetric measurements made with each system alongside the results obtained with a standard diode array. 126 IMRT fields are examined with both EPID dosimetry systems (EPIDose by Sun Nuclear Corporation, Melbourne FL, and Portal Dosimetry by Varian Medical Systems, Palo Alto CA) and th...
In our institution, patient specific quality assurance (QA) for intensity modulated radiation therapy (IMRT) is usually performed by measuring the dose to a point using an ion chamber and by measuring the dose to a plane using film. In order to perform absolute dose comparison measurements using film, an accurate calibration curve should be used. In this paper, we investigated the film response...
IMRT treatment planning with commercial treatment planning systems is a trial-and-error process, based on a series of subjective human decisions. Therefore the quality of the IMRT treatment plans may not be consistent among patients, planners or institutions with different experience. Different plan quality assurance (QA) models have been proposed recently, that could flag suboptimal plans that...
For intensity-modulated radiation therapy, evaluation of the measured dose against the treatment planning calculated dose is essential in the context of patient-specific quality assurance. The complexity of volumetric arc radiotherapy delivery attributed to its dynamic and synchronization nature require new methods and potentially new tools for the quality assurance of such techniques. In the p...
Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in tissue equivalent media. EPIDose (Sun Nu...
The task group (TG) for quality assurance of medical accelerators was constituted by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance and Outcome Improvement Subcommittee. The task group (TG-142) had two main charges. First to update, as needed, recommendations of Table II of the AAPM TG-40 repo...
A spreadsheet based program is presented to perform an independent Monitor Unit (MU) calculation verification for the Quality Assurance (QA) of Intensity Modulated Radiation Therapy (IMRT) using Dynamic MultiLeaf Collimation (DMLC). The computed dose value is compared to the planned dose by calculating the percent dose difference per Intensity Modulated Beam (IMB) and absolute dose difference p...
نمودار تعداد نتایج جستجو در هر سال
با کلیک روی نمودار نتایج را به سال انتشار فیلتر کنید