Commercial breath-alcohol simulators were studied to determine their suitability as calibrating devices and for providing control spocimens in breath-alcohol analysis. The best-fit exponential equation for the air/water partition of ethanol was found to by y = 0.0145e 0.06583x, where x = temperature ~ and y = Ka/wxl03. Temperature, volume, pressure, and alcohol depletion of simulators and their...