Journal: :Medical physics 2008
Philip Vial Peter B Greer Peter Hunt Lyn Oliver Clive Baldock

The purpose of this study was to experimentally quantify the change in response of an amorphous silicon (a-Si) electronic portal imaging device (EPID) to dynamic multileaf collimator (dMLC) beams with varying MLC-transmitted dose components and incorporate the response into a commercial treatment planning system (TPS) EPID prediction model. A combination of uniform intensity dMLC beams and stat...

Ali Asghar Mowlavi, Azam Afzalifar Mohammad Mohammadi, Sara Abdollahi

Introduction: The accuracy of patient positioning plays an important role for radiotherapy tasks. Short and automated Quality Assurance (QA) programs are becoming one of the challenging tasks in Radiotherapy. The current study, investigates the accuracy of treatment couch positioning with four degrees of freedom (4DoF) using a fast and accurate method based on an image acquired...

Journal: :Medical physics 2006
Markus Wendling Robert J W Louwe Leah N McDermott Jan-Jakob Sonke Marcel van Herk Ben J Mijnheer

The use of electronic portal imaging devices (EPIDs) is a promising method for the dosimetric verification of external beam, megavoltage radiation therapy-both pretreatment and in vivo. In this study, a previously developed EPID back-projection algorithm was modified for IMRT techniques and applied to an amorphous silicon EPID. By using this back-projection algorithm, two-dimensional dose distr...

2005
Wendel Dean Renner Kevin Norton Timothy Holmes

A method to convert integrated electronic portal imaging device (EPID) images to fluence for the purpose of reconstructing the dose to a phantom is investigated here for simple open fields. Ultimately, the goal is to develop a method to reconstruct the dose to patients. The EPID images are transformed into incident intensity fluence by spatial filtering with a deconvolution kernel. The kernel u...

2006
Mohammad Mohammadi Eva Bezak Paul Reich

A scanning liquid-filled ionization chamber electronic portal imaging device (SLIC-EPID) and extended dose range (EDR2) films were used to evaluate transmitted dose profiles for homogeneous and inhomogeneous phantoms. Calibrated ionization chamber measurements were used to convert the pixel values acquired from the electronic portal images to dose. Because SLIC-EPID was developed to have a unif...

2014
Stefano Peca Derek W. Brown

The electronic portal imaging device (EPID) has the potential to be used for in vivo dosimetry during radiation therapy as an additional dose delivery check. In this study we have extended a method developed by A. Piermattei and colleagues in 2006 that made use of EPID transit images (acquired during treatment) to calculate dose in the isocenter point. The extension allows calculation of two-di...

2013
Alberto Boriano Francesco Lucio Elisa Calamia Elvio Russi Flavio Marchetto

When using an electronic portal imaging device (EPID) for dosimetric verifications, the calibration of the sensitive area is of paramount importance. Two calibration methods are generally adopted: one, empirical, based on an external reference dosimeter or on multiple narrow beam irradiations, and one based on the EPID response simulation. In this paper we present an alternative approach based ...

Journal: :Medical physics 2004
L N McDermott R J W Louwe J J Sonke M B van Herk B J Mijnheer

The purpose of this study was to investigate the dose-response characteristics, including ghosting effects, of an amorphous silicon-based electronic portal imaging device (a-Si EPID) under clinical conditions. EPID measurements were performed using one prototype and two commercial a-Si detectors on two linear accelerators: one with 4 and 6 MV and the other with 8 and 18 MV x-ray beams. First, t...

2011
Awusi Kavuma Martin Glegg Mohamed Metwaly Garry Currie Alex Elliott

In this study, we use the quadratic calibration method (QCM), in which an EPID image is converted into a matrix of equivalent path lengths (EPLs) and, therefore, exit doses, so as to model doses in conformal and enhanced dynamic wedge (EDW) fields. The QCM involves acquiring series of EPID images at a reference field size for different thicknesses of homogeneous solid water blocks. From these, ...

2014
Shrikant Deshpande Aitang Xing Lois Holloway Peter Metcalfe Philip Vial

The purpose of this study was to investigate the dose response of amorphous silicon (a-Si) electronic portal imaging devices (EPIDs) under different acquisi- tion settings for both open jaw defined fields and segmented intensity-modulated radiation therapy (IMRT) fields. Four different EPIDs were used. Two Siemens and one Elekta plus a standalone Perkin Elmer research EPID. Each was operated wi...