This experimental study shows the validity of Sheridan’s method in determining plasma density in low pressure, weakly magnetized, RF plasmas using ion saturation current data measured by a planar Langmuir probe. The ion density derived from Sheridan’s method which takes into account the sheath expansion around the negatively biased probe tip, presents a good consistency with the electron densit...