In this research, a coated graphite electrode was constructed for determination of Mg2+ in pharmaceutical supplements. The designed sensor demonstrates an ideal Nernstian slope (30.1 mV. Decade-1) over a wide concentration range (1×10-6- 1×10-1 Mol L-1). The selectivity of the sensor was evaluated over 16 different cations by matched potential method and no serious interference was observed fro...