Novel biosensors, based on immobilized creatininase, creatinase and urease are developed, using ISFETs with weak inversion at pH 6-8 and 37.0C. The ISFETs with circuitry, demonstrate a linear relationship of urea and creatinine at the range of 0-200 mM and 0-20 mM, respectively. Preliminary results show that biosensors operating in weak inversion mode can eliminate many of the disadvantages of ...
The Ion Sensitive Field Effect Transistor (ISFET) operation based on polycrystalline silicon thin film transistors is reported. These devices can be fabricated on inexpensive disposable substrates such as glass or plastics and are, therefore, promising candidates for low cost single-use intelligent multisensors. In this work we have developed an extended gate structure with PE-CVD Si3N4 deposit...
The ability to accurately measure real-time pH fluctuations in-vivo could be highly advantageous. Early detection and potential prevention of bacteria colonization of surgical implants can be accomplished by monitoring associated acidosis. However, conventional glass membrane or ion-selective field-effect transistor (ISFET) pH sensing technologies both require a reference electrode which may su...
We offer a comprehensive theory of pH response of a coupled ISFET sensor to show that the maximum achievable response is given by ΔV/ΔpH = 59 mV/pH × α, where 59 mV/pH is the intrinsic Nernst response and α an amplification factor that depends on the geometrical and electrical properties of the sensor and transducer nodes. While the intrinsic Nernst response of an electrolyte/site-binding inter...