Voltammetric Determination of Copper Ion in Water Samples Using a Drop Casted Biopolymer
نویسندگان
چکیده مقاله:
Poly-dopamine, inspired by biopolymers of mussels, is an adhesive catecholamine possessing low cytotoxicity and very good biocompatibility. In the present study, poly-dopamine nanoparticles (PDNP), as a biopolymer, were synthesized and characterized using FTIR spectroscopy, scanning electron microscopy and electrochemical methods. The drop casted PDNPs at the surface of carbon paste electrode (CPE) shows a great tendency to form complexes with Cu(II). This fact was selected for determination of Cu(II) in water. Copper exists broadly in various oxidation states in nature. It is also as one of the most harmful pollutants in the environment due to its non-degradability, high toxicity and bioaccumulation and so determination of it is demanded. X-ray photoelectron spectroscopy (XPS) was applied to clarify the interaction. The influential parameters affecting Cu(II) binding were optimized including adsorption time, pH, and stirring speed. In optimum condition, differential pulse voltammetry was exploited for analytical determination of Cu(II) in the ranges of 1.0 100 µM and 0.05 1.0 µM with the limit of detection 0.04 µM. The sensor ability for Copper determination in water samples was demonstrated.
منابع مشابه
Preconcentration and Determination of Chromium (III) from Sea Water Samples Using Ion Imprinted Activated Carbon
A simple ion imprinted amino-functionalized sorbent was synthesized by coupling activated carbon with iminodiacetic acid, a functional compound for metal chelating, through cyanoric chloride spacer. The resulting sorbent has been characterized using Fourier transform infrared spectroscopy (FT-IR), elemental analysis, thermogravimetric analysis (TGA) and evaluated for the preconcentration and de...
متن کاملPreconcentration and Determination of Chromium (III) from Sea Water Samples Using Ion Imprinted Activated Carbon
A simple ion imprinted amino-functionalized sorbent was synthesized by coupling activated carbon with iminodiacetic acid, a functional compound for metal chelating, through cyanoric chloride spacer. The resulting sorbent has been characterized using Fourier transform infrared spectroscopy (FT-IR), elemental analysis, thermogravimetric analysis (TGA) and evaluated for the preconcentration and de...
متن کاملIon Pair Dispersive Liquid-Liquid Microextraction for the Determination of Trace Amounts of Copper(II) in Soil, Multivitamin Tablet, Tea and Water Samples Using Flame Atomic Absorption Spectrometry
Ion pair dispersive liquid- liquid microextraction (IP-DLLME) method combined with flame atomic absorption spectrometry was proposed for the determination of trace amounts of copper(II). By using pyrocatechol violet as chelating agent and cetyltrimethyl ammonium bromide as an ion pairing agent, the trace amount of copper(II) was extracted in chloroform. The factors influencing the formation cop...
متن کاملA novel voltammetric method for the direct determination of copper in complex environmental samples.
A novel voltammetry with a modified gold electrode for the direct determination of copper in environmental samples, without any pretreatment, is proposed in this paper. A porous disorganized monolayer was formed on the surface of the gold electrode by the self-assembly of mercaptoacetic acid (MAA), which could selectively permeate small molecules. Subtractive square wave anodic stripping voltam...
متن کاملAutomatic simultaneous determination of copper and lead in biological samples by flow injection/stripping voltammetric analysis
An automatic-continuous method for the simultaneous determination of copper and lead based on flow injection analysis (FIA) and stripping voltammetry (SV) is proposed. The method affords the determination of the analytes at the ng/ml level (linear ranges 0.64 to 64.0 ng/ml and 2.1 to 62.2 ng/ml for copper and lead, respectively) with good precision (r.s.d. values smaller than 4%). The selectivi...
متن کاملapplication and construction of carbon paste modified electrodes developed for determination of metal ions in some real samples
ساخت الکترودهاِی اصلاح شده ِیکِی از چالشهاِی همِیشگِی در دانش شیمِی بوِیژه شیمِی تجزیه مِی باشد ،که با در نظر گرفتن سادگِی ساخت، کاربردی بودن و ارزان بودن روش مِی توان به باارزش بودن چنِین سنسورهاِی پِی برد.آنچه که در ادامه آورده شده به ساخت و کاربرد الکترودهاِی اصلاح شده با استفاده از نانو ذرات در اندازه گِیرِی ولتامترِی آهن وکادمِیم اشاره دارد. کار اول اختصاص دارد به ساخت الکترود خمِیر کربن اصلاح شده با لِیگاند داِ...
15 صفحه اولمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 7 شماره 2
صفحات 171- 184
تاریخ انتشار 2020-04-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023