Verification of the Accuracy of the Delivered Dose in Pelvic and Breast Cancer Radiotherapy by in-vivo Semi-Conductor Dosimetry
نویسندگان
چکیده مقاله:
Introduction: Delivering maximum dose to tumor and minimum dose to normal tissues is the most important goal in radiotherapy. According to ICRU, the maximum acceptable uncertainty in the delivered dose compared to the prescribed dose should be lower than 5%, and this is because of the relationship between absorbed dose, tumor control and normal tissue damage. Absorbed dose accuracy is investigated by an in vivo dosimetry method. In this paper, we compared absorbed dose in the tumors of the breast and pelvic region against the calculated dose. The amount of deviations and the factors that cause this deviation in dose delivery to patients and some methods for decreasing them were evaluated. Materials and methods: The entrance and exit doses of 36 pelvic-region cancer patients and 38 breast cancer patients who were treated by cobalt-60 teletherapy were measured using p-type diodes. It should be noted that the transmission method was used to assess the dose at isocenter. Two ionization chambers (0.6 cc and 0.3 cc) were used for calibration and determination of the correction coefficients in water and slab phantoms. Deviations between calculated and measured doses of entrance, exit and midline points were calculated and the results were shown using histograms. Results: The average and standard deviation for entrance, exit and midline points for pelvis cancer were assessed to be about 0.10%, -1.86% and -1.35% for mean deviation and 5.03%, 7.32% and 5.86% for standard deviation, respectively. The corresponding data for breast cancer were 0.78%, 5.29% and 3.59% for mean deviation and 5.97%, 10.23% and 9.86%, respectively. There was no significant difference between the calculated and measured doses (p > 0.1), except exit dose in breast cancer (p < 0.05). The temperature and angle of incidence correction factors were neglected due to their less than 1% deviations. Discussion and Conclusions: Some error sources are patient setup error, patient motion and dose calculation algorithm error (due to ignoring inhomogeneity and patient curvature). As no significant deviations were found in midline dose, the method used has an acceptable accuracy. In vivo dosimetry can perform a basic role in the quality control of radiotherapy departments.
منابع مشابه
verification of the accuracy of the delivered dose in pelvic and breast cancer radiotherapy by in-vivo semi-conductor dosimetry
introduction: delivering maximum dose to tumor and minimum dose to normal tissues is the most important goal in radiotherapy. according to icru, the maximum acceptable uncertainty in the delivered dose compared to the prescribed dose should be lower than 5%, and this is because of the relationship between absorbed dose, tumor control and normal tissue damage. absorbed dose accuracy is investiga...
متن کاملVerification of the Accuracy of the Delivered Dose in Brain Tumors by in Vivo Dosimetry Using Diode Detectors
Introduction: During radiotherapy, high accuracy in the dose delivery is required because there is a strong relationship between the absorbed dose, local tumor control and particularly the normal tissue damage. In many institutions, in vivo dosimetry using diodes is performed to check the actual dose delivered. In general, the uncertainty in the dose delivered should fall within ± 5% of the...
متن کاملthe study of aaag repeat polymorphism in promoter of errg gene and its association with the risk of breast cancer in isfahan region
چکیده: سرطان پستان دومین عامل مرگ مرتبط با سرطان در خانم ها است. از آنجا که سرطان پستان یک تومور وابسته به هورمون است، می تواند توسط وضعیت هورمون های استروئیدی شامل استروژن و پروژسترون تنظیم شود. استروژن نقش مهمی در توسعه و پیشرفت سرطان پستان ایفا می کند و تاثیر خود را روی بیان ژن های هدف از طریق گیرنده های استروژن اعمال می کند. اما گروه دیگری از گیرنده های هسته ای به نام گیرنده های مرتبط به ا...
15 صفحه اولverification of the accuracy of the delivered dose in brain tumors by in vivo dosimetry using diode detectors
introduction: during radiotherapy, high accuracy in the dose delivery is required because there is a strong relationship between the absorbed dose, local tumor control and particularly the normal tissue damage. in many institutions, in vivo dosimetry using diodes is performed to check the actual dose delivered. in general, the uncertainty in the dose delivered should fall within ± 5% of the...
متن کاملstudy of cohesive devices in the textbook of english for the students of apsychology by rastegarpour
this study investigates the cohesive devices used in the textbook of english for the students of psychology. the research questions and hypotheses in the present study are based on what frequency and distribution of grammatical and lexical cohesive devices are. then, to answer the questions all grammatical and lexical cohesive devices in reading comprehension passages from 6 units of 21units th...
a study of the fifth child and ben in the world by doris lessing in the light of julia kristevas psychoanalytic concepts
این مطالعه به بررسی عوامل روانشناختی کریستوادردو رمان دوربس لسینگ،فرزندبنجم و بن دردنیای واقعی می بردازد.موفقیت یا شکست کاراکترهادر تکمیل شکست تتیزی یه کمک بدر وهم از مهم ترین دغدغه محقق می باشد.به بررسی تحلیل روانشناختی تمامی کاراکترها خصوصا بن برداخته و به دنبال نشانه هایی از جامعه شیشه ای کریستوا می باشد.
15 صفحه اولمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 6 شماره Issue 3,4
صفحات 38- 45
تاریخ انتشار 2009-12-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023