Photon Dosimetry Quality Audit of Radiotherapy Centers in Iran: SSDL's On-Site Visit Program During 2007-2017
نویسندگان
چکیده مقاله:
Introduction: A dosimetric audit is a quality assurance process (QA) that allows the accuracy of dosimetric and geometric precision to accurately control the delivery of doses. They can be used to evaluate the accuracy of radiation delivery at different radiotherapy centers. Differences in prescriptive doses, depending on the severity of the differences, have potential consequences not so important, severe and even fatal. Measuring the absolute output of linear accelerators (LINACs) under reference conditions is a standard method for verifying the accuracy of dose delivery. In this study, Radiotherapy centers in Iran were audited by the secondary standard dosimetry laboratory (SSDL). The absolute photon output of the accelerators and some of the relative factors such as wedge factor, shield tray factor, total scatter factor was measured using a Farmer chamber and the results were compared with the reported values by the radiotherapy physicist of centers. The differences between SSDL and the inspected center were reported. Materials and Methods: Measurements were performed at 178 on-site visits from 61 radiotherapy centers (including 98 LINACs) between March 2007 and October 2017. Some participating centers were visited more than once. Photon absolute dose measurements were carried out according to TRS-398 protocol using Farmer ionization chamber (TW30013). Action levels were determined including ±3% for absolute dose and ±2% for relative factors. Results: Absolute dose differences between SSDL and local centers were in the range of -3.5% up to 4%. The differences in the values of the wedge factor between the amounts measured by SSDL and the local centers were -3.5% up to 3.36%. The minimum and maximum differences in the total scatter factor were -3.4% up to 3.7%. The differences in the obtained tray factor values were -1.4% up to 2.5%. The difference between SSDL and radiotherapy centers at the reported absolute dose values in 11 beam measurements was greater than the permissible limits. Moreover, 299 beam measurements were within the tolerance. Conclusion: Differences in measurements between SSDL and radiotherapy centers could be due to the following: I. Lack of qualified medical physicists in some radiotherapy centers, II. Error in calculating beam quality factor (kQ) for some chambers, III. Mistake in the report of the reference depth and reference source to surface distance (SSD).
منابع مشابه
In vivo dosimetry of intraoral stent using TLD during external photon beam radiotherapy of oral cavity
Introduction: Individual oral stent is a mouth-opening device that used in head and neck cancer radiotherapy with the intention of decreasing radiation dose to health tissues. The aim of this study was to check the dose delivered to both the hard palate and tongue when patient uses the stent during radiotherapy of tongue carcinoma. Methods and materials<stron...
متن کاملthe impact of e-readiness on ec success in public sector in iran the impact of e-readiness on ec success in public sector in iran
acknowledge the importance of e-commerce to their countries and to survival of their businesses and in creating and encouraging an atmosphere for the wide adoption and success of e-commerce in the long term. the investment for implementing e-commerce in the public sector is one of the areas which is focused in government‘s action plan for cross-disciplinary it development and e-readiness in go...
a swot analysis of the english program of a bilingual school in iran
با توجه به جایگاه زبان انگلیسی به عنوان زبانی بین المللی و با در نظر گرفتن این واقعیت که دولت ها و مسئولان آموزش و پرورش در سراسر جهان در حال حاضر احساس نیاز به ایجاد موقعیتی برای کودکان جهت یاد گیری زبان انگلیسی درسنین پایین در مدارس دو زبانه می کنند، تحقیق حاضر با استفاده از مدل swot (قوت ها، ضعف ها، فرصتها و تهدیدها) سعی در ارزیابی مدرسه ای دو زبانه در ایران را دارد. جهت انجام این تحقیق در م...
15 صفحه اولValidating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service.
In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage p...
متن کاملthe effect of e-64 on the developmental competence of sheep cocs during in vitro maturation
in the present study, the effect of e-64 at different concentrations (0.5, 1, 5 and 10 µm) added to (1) the ivm medium on oocyte nuclear maturation and developmental competence of ovine oocytes, and (2) to the ivc medium on embryonic development of ovine embryos were investigated.
Implementation of an intraoperative electron radiotherapy in vivo dosimetry program
BACKGROUND Intraoperative electron radiotherapy (IOERT) is a highly selective radiotherapy technique which aims to treat restricted anatomic volumes during oncological surgery and is now the subject of intense re-evaluation. In vivo dosimetry has been recommended for IOERT and has been identified as a risk-reduction intervention in the context of an IOERT risk analysis. Despite reports of fruit...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 15 شماره Special Issue-12th. Iranian Congress of Medical Physics
صفحات 121- 121
تاریخ انتشار 2018-12-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023