Patient-Specific Radiation Dose and Cancer Risk in Computed Tomography Examinations in Ondo, Nigeria
نویسندگان
چکیده مقاله:
Introduction: The dose in computed tomography (CT) often approach or exceed the optimum levels, thereby increasing the probability of cancer induction. With wide application of this diagnostic test, it is expedient to determine the effective dose (ED) for each patient to estimate their cancer risk. This study was conducted to investigate the patient-specific dose (PSD) and cancer risk in CT examinations in Ondo, Nigeria. Material and Methods: The study was conducted on 160 patients undergoing eight most common types of CT examinations performed at the center, from December 5, 2015 to February 28, 2016. Body mass index (BMI) was calculated for each patient using demographic data, PSD was determined and estimation of the lifetime attributable risk (LAR) of cancer was accomplished using the Biologic Effects of Ionizing Radiation VII (2006) report phase 2 models. Results: From the results, radiation doses varied significantly within and between the types of CT examinations. The mean ED was 5.88±3.75 mSv in a range of 0.78-19.00 mSv. The mean PSD was 0.274±0.229 mSv/kgm-2 in a range of 0.024-1.555 mSv/kgm-2 and the mean LAR of cancer incidence was 0.04861±0.03996 Sv-1 in a range of 0.00004-0.21942 Sv-1. Conclusion: ED and PSD varied within and across the CT examinations. In this regard, the coefficients of variation of ED for abdominal, cranial, craniocervical, abdomen/pelvis, thoracic, thoracoabdominal, cervical spine, and pelvis were 5.7%, 6.6%, 3.9%, 8.9%,3.7%, 6.0%, 44.7%, and 19.2%, respectively. Accordingly, the coefficients of variation of PSD were 9.0%, 7.9%, 7.0%, 10.1%,5.6%,23.8%, 47.7%, and 14.2%.
منابع مشابه
Diagnostic Reference Level of Computed Tomography Examinations and Need for Dose Optimization in Ondo State, Nigeria
Introduction: The present study was conducted to obtain State diagnostic reference levels (DRLs) of five routine computed tomography (CT) examinations from two CT centers in Ondo State and to identify factors responsible for dose variation and escalation in these CT centers. Material and Methods: Acquisition parameters and CT dose indices wer...
متن کاملPatient effective dose estimation for routine computed tomography examinations in Iran
Background: The present study intended to determine and report patient effective dose on the basis of patients and exposure data. Materials and methods: A nationwide computed tomography (CT) survey was provided as a report of patient doses in 2015-2016. Scan details were collected for nearly 2,000 adults and children in four age groups subjected to CT examinations. From total 565 CT scanners in...
متن کاملDiagnostic Reference Levels for Computed Tomography Examinations in Iran: A Nationwide Radiation Dose Survey
Introduction: International Commission on Radiological Protection introduced three basic principles of radiation protection, namely justification, optimization, and dose limit. Medical exposure has no dose limits, and generally, diagnostic reference levels are used as a tool for optimization of patient protection. Material and Methods: Dosimetry was performed on 20 CT scanners located in 14 cit...
متن کاملRadiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer.
BACKGROUND Use of computed tomography (CT) for diagnostic evaluation has increased dramatically over the past 2 decades. Even though CT is associated with substantially higher radiation exposure than conventional radiography, typical doses are not known. We sought to estimate the radiation dose associated with common CT studies in clinical practice and quantify the potential cancer risk associa...
متن کاملEffect of Phantom Size and Tube Voltage on the Size-Conversion Factor for Patient Dose Estimation in Computed Tomography Examinations
Introduction: This study aimed to establish the conversion factors to normalize the output dose of volumetric computed tomography dose index (CTDIvol) to the patient dose (i.e. size-specific dose estimate (SSDE)) for various phantom diameters and tube voltages. Material and Methods: In-house cylindrical acrylic phantoms with physical diameter...
متن کاملEstimating the Radiation-Induced Cancer Risks in Pediatric Computed Tomography
Introduction One of the central questions in radiological protection is the magnitude of the risks from low doses of radiation, related to the justification and optimization of the diagnostic medical exposures. Therefore, the aim of this study was to estimate the cancer incidence and mortality risks in children of different ages, sizes, and ethnicities undergoing computed tomography examination...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 16 شماره 1
صفحات 85- 90
تاریخ انتشار 2019-01-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023