Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry
نویسندگان
چکیده مقاله:
Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clinac iX 10 MV photon beam for square small field size (1×1, 2×2, 3×3, 4×4 and 5×5 cm2) using Monte Carlo (MC) simulation. Materials and Methods: The commissioning process for this linac, has been conducted for field size 6×6, 10×10 and 20×20 cm2 by comparing the measurement and MC simulation data. Head linac simulation was performed with BEAMnrc and dose calculation with DOSXYZnrc. The phase space (phsp) data from BEAMnrc was analyzed using BEAMDP to get the particles information in scoring plane. Results and Discussion: The scatter angle of particles depends on the field size. This factor affects the penumbra width in water phantom. On the other hand, PDD data show that the depth of maximum dose and penumbra width in small field shifted correspond with the number of scatter particle. The difference of relative output factor between measurement and MC results were found less than 2%. However, the 2% difference was still acceptable in photon beam dosimetry. Conclusion: From this simulation, the electron contamination give contribution in surface dose of water phantom about 13.0581% and less than 1% for field size 10×10 cm2 and small field size, respectively.
منابع مشابه
Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation
Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...
متن کاملsmall photon field dosimetry using ebt2 gafchromic film and monte carlo simulation
background: small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (imrt) and stereotactic radiosurgery (srs) treatments. accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. hence the ebt2 (external beam therapy) gafchromic film was used for dosimetry of sm...
متن کاملmonte carlo study of fetal dosimetry parameters for 6 mv photon beam
background: because of the adverse effects of ionizing radiation on fetuses, prior to radiotherapy of pregnant patients, fetal dose should be estimated. fetal dose has been studied by several authors in different depths in phantoms with various abdomen thicknesses (at). in this study, the effect of maternal at and depth in fetal dosimetry was investigated, using peripheral dose (pd) distributio...
متن کاملmonte carlo simulation of varian clinac 2100c electron beams
background: the purpose of this study was to investigate the application of the monte carlo technique to calculate and analysis of dosimetric parameters for electron beams used in radiotherapy. this technique is based on statistical method and has a powerful role in different radiotherapy aspects. materials and methods: the simulated medical linear accelerator was the varian clinac 2100c. the e...
متن کاملMonte Carlo Study of Fetal Dosimetry Parameters for 6 MV Photon Beam
Because of the adverse effects of ionizing radiation on fetuses, prior to radiotherapy of pregnant patients, fetal dose should be estimated. Fetal dose has been studied by several authors in different depths in phantoms with various abdomen thicknesses (ATs). In this study, the effect of maternal AT and depth in fetal dosimetry was investigated, using peripheral dose (PD) distribution evaluatio...
متن کاملOn the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry.
Monte Carlo (MC) calculated detector-specific output correction factors for small photon beam dosimetry are commonly used in clinical practice. The technique, with a geometry description based on manufacturer blueprints, offers certain advantages over experimentally determined values but is not free of weaknesses. Independent MC calculations of output correction factors for a PTW-60019 micro-di...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 15 شماره None
صفحات 275- 282
تاریخ انتشار 2017-07
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023