Measuring the Skyshine from Linear accelerator in Rajaee Oncology Hospital
نویسندگان
چکیده مقاله:
Introduction: Radiation therapy is using the ionizing radiation to eliminate or minimize cancerous tissues. This method has been very much considered in recent years. Because of the high energy of the photons produced by the accelerator, it is necessary to protect the walls, roof and the floor of the treatment rooms using the concrete and leaden materials. One of these harmful radiations produced in radiation therapy centers is the Skyshine that should be considered in designing the shieldings. Skyshine rays are the primary beams that pass through the ceiling in the vertical direction and scatter in the upper atmosphere above the building and return to the ground; so they transfer radiation dose to a general area outside the radiotherapy room. This study aimed to evaluate and determine the levels of doses resulted from skyshine radiation in Shahid Rajaee Radiation Center of Babolsar in Northern Province of Iran and compare it with the values of international standards. Materials and Methods: In this survey dose measurement has been done with LS01 ion chamber (1000 cc) and Unidose E electrometer produced by PTW company in 6mv Elekta compact accelerator (made in Germany) with exposing 400MUs and field size of 40*40. The distances of dose measurement were 300, 150, 100, 50 and 25 cm from the external wall of the treatment room and 2.10cm height from the ground. Temperature, pressure and correction factors of electrometer and dosimeter has been considered in results and total dose of skyshine rays based on reformed reading of electrometer acquired for 50 weeks,6 days and 8 working hours. Results: The dose rate at 25, 50, 100, 150 and 300 cm distances from the external wall of the treatment room were 1.017, 0.79, 0.595, 0.436, 0.275 µsv/h respectively. Also, the standard deviations were 0.012, 0.007, 0.015, 0.020, and 0.032. Conclusion: The obtained results were compared with other studies and the annual average worldwide rate; the result was compatible with most studies and was less than the average worldwide= 2.4 mSv/y.
منابع مشابه
Accelerator boundary doses and skyshine.
Beginning this year, AEC contractors are being required to submit dose evaluations as low as 1.7 mrem per year to the surrounding public from their installations. This small incremental dose exists within the framework of a natural background of about 130 mrem per year. Some of the pitfalls in estimating these small doses, especially with regard to skyshine, are discussed. It is shown that, for...
متن کاملMothers’ Satisfaction Rate from Hospital Cares in Hematology- Oncology Ward
Background Satisfaction evaluation is a good way to assess hospital conditions. In Health Care System, parentscan be also as children's main supporters, thus they may act as patient's viewpoints' representatives.This study aimed to evaluate mother’s satisfaction of hospital care in hematology – oncology ward in Dr Sheikh hospital. Materials and Methods A Cross-sectional descriptive analytic stu...
متن کاملThe Effect of Field Size and Distance from the Field Center on Neutron Contamination in Medical Linear Accelerator
Objective: Using Megavoltage photons generated by medical linear accelerator is a common modality for the treatment of malignant. The crucial problem for using photon beams >8MV is the photoneutron yields that increase the risk of secondary cancer that treated with high-energy photon beams. The contaminated neutrons produced in different components of the accelerator head and rely on many param...
متن کاملThe Effect of Field Size and Distance from the Field Center on Neutron Contamination in Medical Linear Accelerator
Objective: Using Megavoltage photons generated by medical linear accelerator is a common modality for the treatment of malignant. The crucial problem for using photon beams >8MV is the photoneutron yields that increase the risk of secondary cancer that treated with high-energy photon beams. The contaminated neutrons produced in different components of the accelerator head and rely on many param...
متن کاملAssessment of skyshine photon dose rates from 9 and 18 MV medical linear accelerators
Background: Skyshine describes the radiation scattered by the atmosphere above a LINAC facility to a point on the ground. The aim of this study was to measure the skyshine photon dose rates from two different (9 MV and 18 MV) medical linear accelerators. Materials and Methods: The photon beam was directed upward (180° gantry position), with a maximum photon field size (40 × 40 cm2) at the isoce...
متن کاملSkyshine photon doses from 6 and 10 MV medical linear accelerators
The skyshine radiation phenomenon consists of the scattering of primary photon beams in the atmosphere above the roof of a medical linear accelerator facility, generating an additional dose at ground level in the vicinity of the treatment room. Thus, with respect to radioprotection, this situation plays an important role when the roof is designed with little shielding and there are buildings ne...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 15 شماره Special Issue-12th. Iranian Congress of Medical Physics
صفحات 71- 71
تاریخ انتشار 2018-12-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023