Investigation of Neutron Contamination of Flattening Filter and Flattening Filter-Free 10-MV Photon Beams in Elekta InfinityTM Accelerator

نویسندگان

  • Fauzia Puspa Lestari Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung
  • Freddy Haryanto Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung
  • Indra Budiansah Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung
  • Mohamad Rhani Department of Radiation Oncology, Concord International Hospital, Singapore/ Singapore
  • Sitti Yani Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung
چکیده مقاله:

Introduction: This study aimed to measure the neutron contamination of flattening filter (FF) and flattening filter-free (FFF) 10-MV photon beams delivered by the Elekta InfinityTM accelerator. Material and Methods: The photoneutron spectrum produced by the Linac head was evaluated using a Monte Carlo (MC) simulation. The geometry and composition of the head Linac material were modelled based on information obtained from the manufacturer. In this simulation, MC N-Particle Transport Code software (MCNP6) was utilized to model the Linac head and simulate the particle transport. Evaluation of neutron contamination was carried out for the Linac with FF and without it (i.e., FFF). In this regard, the FFF beam was built by removing the FF from the Linac components. The scoring plane, as the neutron spectra calculation area for FF and FFF beams, was placed 99 cm from the target. Results: The neutron type produced by the head Linac Elekta InfinityTM 10-MV photon mode was mostly thermal and fast. Although there were differences in the neutron intensity of FF and FFF beams, the type of neutrons produced by these two modes had the same energy. Based on the photoneutron reaction energy threshold, it can be concluded that the neutrons produced from the head Linac were the result of photoneutron interactions of high-energy photons with molybdenum-96 and tungsten-184 isotopes. Conclusion: The photoneutron quantity did not change for FF and FFF beams; however, a larger quantity of neutrons was produced in the FF beam.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Surface dose variations in 6 and 10 MV flattened and flattening filter‐free (FFF) photon beams

As the use of linear accelerators operating in flattening filter-free (FFF) modes becomes more widespread, it is important to have an understanding of the surface doses delivered to patients with these beams. Flattening filter removal alters the beam quality and relative contributions of low-energy X-rays and contamination electrons in the beam. Having dosimetric data to describe the surface do...

متن کامل

Dosimetric influence of Flattening Filter (FF) and Flattening Filter Free (FFF) 6 and 10 MV photon beams on Volumetric Modulated Arc Therapy (VMAT) planning in case of prostate carcinoma

Background: In the treatment of prostate cancer, radiotherapy is the potential to increase second primary cancers such as bladder and rectal cancers. The reasons for this potential are more monitor units (MUs), therefore a larger total body dose because of leakage radiation, a bigger volume of normal tissue is exposed to lower radiation doses. This study was designed to compare the integral dos...

متن کامل

Beam Characterization of 10-MV Photon Beam from Medical Linear Accelerator without Flattening Filter

AIM This work investigated the dosimetric properties of a 10-MV photon beam emitted from a medical linear accelerator (linac) with no flattening filter (FF). The aim of this study is to analyze the radiation fluence and energy emitted from the flattening filter free (FFF) linac using Monte Carlo (MC) simulations. MATERIALS AND METHODS The FFF linac was created by removing the FF from a linac ...

متن کامل

Dosimetric properties of photon beams from a flattening filter free clinical accelerator.

Basic dosimetric properties of 6 MV and 18 MV photon beams from a Varian Clinac 21EX accelerator operating without the flattening filter have been measured. These include dose rate data, depth dose dependencies and lateral profiles in a water phantom, total scatter factors and transmission factors of a multileaf collimator. The data are reviewed and compared with measurements for the flattened ...

متن کامل

Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy

Background: This study aims to investigate radiation beam geometry of Cyberknife beam and change in dosimetric characteristics of six megavoltage (6MV) flattening filter free (FFF) beam after passing through high density cadmium free compensator alloy.Material and Methods: In this experimental study, changes in FFF beam dosimetric characteristics after passing through compensator alloy wa...

متن کامل

Effect of dose rate on antitumor activity in hypoxic cells by using flattening filter free beams

Background: Recently, the Flattening Filter Free (FFF) beams, which allow a dose rate increase of up to four times compared to the normal dose rate, have been incorporated into radiation therapy machines. The aim of this study is to evaluate antitumor activity in hypoxic cells irradiated with different dose rates using FFF beams and to identify the casual mechanism for cellular damage during ir...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید