In Vivo Dosimetry Using a Flat Surface Sun Nuclear Corporation Diode in 60co Beams for Some Radiotherapy Treatments in Ghana
نویسندگان
چکیده مقاله:
Introduction: One of the useful standard quality assurance techniques in radiation therapy is monitoring entrance doses in in-vivo dosimetry. An overall tolerance limit of 5% of the absorbed radiation dose has been recommended by the International Commission of Radiological Units. The implementation of an in vivo dosimetry still remains as a challenge to clinical medical physicists. As a result, the practice of constant monitoring of patients undergoing radiation therapy in most of the radiotherapy departments in Africa has not been given much attention. The study aimed at the evaluation of in-vivo entrance dosimetry using diodes to verify the accuracy of the radiation delivered to patients, compared to prescribed doses. Material and Methods: In this paper, a protocol for in vivo dosimetry using a two flat surface Sun Nuclear Corporation diode in a radiotherapy department has been implemented in equinox Cobalt 60 beams. A water phantom calibrated was performed using the International Atomic Energy Agency standards (TRS 398). Calibration coefficients were determined with diodes using a Perspex phantom to derive correction factors. A total number of 137 patients’ doses were measured with the diodes during the treatment of 4 different sites. Results: The average deviation between the measured and expected entrance dose performed by the phantom studies was 5% (0.34±1.8%) in almost all cases. Conclusion: The developed protocol in this study indicates that in vivo dosimetry using silicon diodes is reliable, which can be adopted as a universal quality assurance tool in the radiotherapy departments. Moreover, measurements with diodes can be acquired online which produces an instant readout and is relatively cheaper as compared to the ion chamber.
منابع مشابه
Skin in vivo Dosimetry in Radiotherapy
Introduction: Due to the prevalence of skin problems in patients after radiotherapy, skin dose measuring is importance. Content: Skin in vivo dosimetry means measuring the patient's (or phantom) skin dose during radiotherapy. According to the ICRP 59, the dose at the depth of 0.07 mm is known as a skin dose. The most radiosensitive epidermis cells are located...
متن کاملIn vivo dosimetry using a single diode for megavoltage photon beam radiotherapy: Implementation and response characterization
The AAPM Task Group 40 reported that in vivo dosimetry can be used to identify major deviations in treatment delivery in radiation therapy. In this paper, we investigate the feasibility of using one single diode to perform in vivo dosimetry in the entire radiotherapeutic energy range regardless of its intrinsic buildup material. The only requirement on diode selection would be to choose a diode...
متن کاملpassivity in waiting for godot and endgame: a psychoanalytic reading
this study intends to investigate samuel beckett’s waiting for godot and endgame under the lacanian psychoanalysis. it begins by explaining the most important concepts of lacanian psychoanalysis. the beckettian characters are studied regarding their state of unconscious, and not the state of consciousness as is common in most beckett studies. according to lacan, language plays the sole role in ...
Retrospective dosimetry using fingernail in nuclear industries: A review paper
Generally, doses greater than 2Gy can produce acute biological effects in humans. For this reason, to measure and record the cumulative dose, especially in radiation users, the need for a personal dosimeter has been strongly felt by the researchers. The purpose of this study is to introduce retrospective dosimetric techniques, in particular the use of nails as a natural tool for determining cum...
متن کاملIn vivo dosimetry of intraoral stent using TLD during external photon beam radiotherapy of oral cavity
Introduction: Individual oral stent is a mouth-opening device that used in head and neck cancer radiotherapy with the intention of decreasing radiation dose to health tissues. The aim of this study was to check the dose delivered to both the hard palate and tongue when patient uses the stent during radiotherapy of tongue carcinoma. Methods and materials<stron...
متن کاملمنابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
ذخیره در منابع من قبلا به منابع من ذحیره شده{@ msg_add @}
عنوان ژورنال
دوره 16 شماره 5
صفحات 329- 335
تاریخ انتشار 2019-09-01
با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.
کلمات کلیدی
میزبانی شده توسط پلتفرم ابری doprax.com
copyright © 2015-2023