Energy and field size dependence of a silicon diode designed for small-field dosimetry

نویسندگان

  • Mehran Yarahmadi Department of Medical Physics, Faculty of Medicine, Kurdistan University of Medical Sciences, Sanandaj, Iran
  • Otto A. Sauer Department of Radiation Oncology, University of Wuurzburg, Josef-Schneider-Strasse 11, 97080 Wuurzburg, Germany
  • Sonja Wegener Department of Radiation Oncology, University of Wuurzburg, Josef-Schneider-Strasse 11, 97080 Wuurzburg, Germany
چکیده مقاله:

Purpose: To investigate the energy dependence/spectral sensitivity of silicon diodes designed for small-field dosimetry and obtain response factors (RFs) for arbitrary photon spectra using Monte Carlo (MC) simulations.   Methods: The EGSnrc user-code DOSRZnrc was used to calculate the dose deposition in water and in the active volume of a stereotactic diode field detector (SFD). Then, the RFs of the SFD were calculated for several circular field sizes and energies at 5 cm depth in water. Several low-energy photon spectra (mean energy 55 to 200 keV), as well as Co-60 radiation (mean energy 1.25 MeV) and a 6 MV Elekta Synergy beam (mean energy 2.9 MeV), in 10×10 cm2 field size were used to validate the MC calculations, using a simple beam model. The RFs of the SFD detector for a 6 MV Elekta Synergy linac photon beam in different field sizes were calculated. These were also measured with EBT3 Gafchromic film and the SFD detector.   Results: For the reference field size, differences between measured and calculated RFs were less than 5% at mean energies below 1 MeV and less than 1% at energies above 1 MeV. The calculated RFs for a 6 MV Elekta Synergy linac photon beam as a function of different field sizes showed a good agreement between the measurements and previously reported results. This agreement was within 2% for all considered field sizes.   Conclusion: While at high photon energies, the change of response of the SFD is marginal, whereas it is extreme at low energies. Therefore, it is desirable to benchmark response calculations also in the low energy domain. Our results, with a simple beam model and geometry, indicate that a validation of the simulations by experimental results is achievable. The present work provides a comprehensive table that can be used to calculate SFD detector response factors depending on both, field size and photon energy.

برای دانلود باید عضویت طلایی داشته باشید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry

Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...

متن کامل

simulation and experimental studies for prediction mineral scale formation in oil field during mixing of injection and formation water

abstract: mineral scaling in oil and gas production equipment is one of the most important problem that occurs while water injection and it has been recognized to be a major operational problem. the incompatibility between injected and formation waters may result in inorganic scale precipitation in the equipment and reservoir and then reduction of oil production rate and water injection rate. ...

Monte Carlo-based diode design for correction-less small field dosimetry.

Due to their small collecting volume, diodes are commonly used in small field dosimetry. However, the relative sensitivity of a diode increases with decreasing small field size. Conversely, small air gaps have been shown to cause a significant decrease in the sensitivity of a detector as the field size is decreased. Therefore, this study uses Monte Carlo simulations to look at introducing air u...

متن کامل

Validation of a prototype DiodeAir for small field dosimetry.

Standard commercial diode detectors over-respond within small radiation fields, an effect largely attributable to the relatively high mass-density of silicon. However, Monte Carlo studies can be used to optimise dosimeter designs and have demonstrated that 'mass-density compensation'-for example, introducing a low-density air-gap upstream of a diode's high-density silicon volume-can substantial...

متن کامل

Dosimetry for Small, Irregular and Rectangular Field Size for Electron Beam Therapy

Purpose: This study was to investigate the effect of irregularity in shape of field size of electron beam on the percentage depth dose and the output factor. Study the variety of the Percentage Depth Dose (PDD) and the output factors for rectangular and small field sizes when one of field dimensions less than the Practical Range (Rp) of the electron energy used. Compare the measured output fact...

متن کامل

Small photon field dosimetry using EBT2 Gafchromic film and Monte Carlo simulation

Background: Small photon fields are increasingly used in modern radiotherapy especially in intensity modulated radiation therapy (IMRT) and stereotactic radiosurgery (SRS) treatments. Accurate beam profile and central axis depth doses measurements of such beams are complicated due to the electron disequilibrium. Hence the EBT2 (external beam therapy) Gafchromic film was used for dosimetry of sm...

متن کامل

منابع من

با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

ذخیره در منابع من قبلا به منابع من ذحیره شده

{@ msg_add @}


عنوان ژورنال

دوره 15  شماره Special Issue-12th. Iranian Congress of Medical Physics

صفحات  409- 409

تاریخ انتشار 2018-12-01

با دنبال کردن یک ژورنال هنگامی که شماره جدید این ژورنال منتشر می شود به شما از طریق ایمیل اطلاع داده می شود.

میزبانی شده توسط پلتفرم ابری doprax.com

copyright © 2015-2023